Preliminary fabrication and characterization of electron beam melted Ti–6Al–4V customized dental implant
المؤلفون المشاركون
Divakar, Darshan Devang
Hashim, Muhammad Ibrahim
al-Kharif, Abd al-Aziz Abd Allah
Muhammad, Ashfaq
Kotha, Sunil Babu
Celur, Sree Lalita
Rahman, Ihtisham Ur
Vallittu, Pekka K.
Ramakrishnaiah, Ravikumar
المصدر
Saudi Journal of Biological Sciences
العدد
المجلد 24، العدد 4 (31 مايو/أيار 2017)، ص ص. 787-796، 10ص.
الناشر
تاريخ النشر
2017-05-31
دولة النشر
السعودية
عدد الصفحات
10
التخصصات الرئيسية
الموضوعات
الملخص EN
The current study was aimed to fabricate customized root form dental implant using additive manufacturing technique for the replacement of missing teeth.
The root form dental implant was designed using GeomagicTM and MagicsTM, the designed implant was directly manufactured by layering technique using ARCAM A2TM electron beam melting system by employing medical grade Ti–6Al–4V alloy powder.
Furthermore, the fabricated implant was characterized in terms of certain clinically important parameters such as surface microstructure, surface topography, chemical purity and internal porosity.
Results confirmed that, fabrication of customized dental implants using additive rapid manufacturing technology offers an attractive method to produce extremely pure form of customized titanium dental implants, the rough and porous surface texture obtained is expected to provide better initial implant stabilization and superior osseointegration.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Ramakrishnaiah, Ravikumar& al-Kharif, Abd al-Aziz Abd Allah& Muhammad, Ashfaq& Divakar, Darshan Devang& Kotha, Sunil Babu& Celur, Sree Lalita…[et al.]. 2017. Preliminary fabrication and characterization of electron beam melted Ti–6Al–4V customized dental implant. Saudi Journal of Biological Sciences،Vol. 24, no. 4, pp.787-796.
https://search.emarefa.net/detail/BIM-782308
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Ramakrishnaiah, Ravikumar…[et al.]. Preliminary fabrication and characterization of electron beam melted Ti–6Al–4V customized dental implant. Saudi Journal of Biological Sciences Vol. 24, no. 4 (May. 2017), pp.787-796.
https://search.emarefa.net/detail/BIM-782308
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Ramakrishnaiah, Ravikumar& al-Kharif, Abd al-Aziz Abd Allah& Muhammad, Ashfaq& Divakar, Darshan Devang& Kotha, Sunil Babu& Celur, Sree Lalita…[et al.]. Preliminary fabrication and characterization of electron beam melted Ti–6Al–4V customized dental implant. Saudi Journal of Biological Sciences. 2017. Vol. 24, no. 4, pp.787-796.
https://search.emarefa.net/detail/BIM-782308
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references : p. 795-796
رقم السجل
BIM-782308
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر