On the Optimum Sampling Time for the Measurement of Pollutants in Ambient Air
المؤلفون المشاركون
Hood, Daniel
Brown, Richard J. C.
Brown, Andrew S.
المصدر
Journal of Analytical Methods in Chemistry
العدد
المجلد 2008، العدد 2008 (31 ديسمبر/كانون الأول 2008)، ص ص. 1-6، 6ص.
الناشر
Hindawi Publishing Corporation
تاريخ النشر
2008-01-03
دولة النشر
مصر
عدد الصفحات
6
التخصصات الرئيسية
الملخص EN
No measurement result of any kind is complete without an associated statement of uncertainty.
In general, the aim is to minimise this measurement uncertainty to the extent that the measurement made is fit for purpose, or the uncertainty meets an acceptable level such as a ‘‘data quality objective’’.
This paper demonstrates how the sampling time used to measure a pollutant in ambient air over a given time period, particularly with automatic instruments, can affect the uncertainty of the measurement result.
It is shown that the uncertainty of the overall result depends critically on how the precision characteristics of the measuring instrumentation vary with analyte abundance.
An example of the determination of these characteristics for mercury vapour measurement and the conclusions that may be drawn about optimum sampling times to minimise uncertainty are presented.
For situations where uncertainty is minimised by the use of long sampling periods, the implications of this strategy on the ‘‘information richness’’ of the data and on the detection of short-term episodes, as well as peak concentrations, are discussed.
نمط استشهاد جمعية علماء النفس الأمريكية (APA)
Brown, Richard J. C.& Hood, Daniel& Brown, Andrew S.. 2008. On the Optimum Sampling Time for the Measurement of Pollutants in Ambient Air. Journal of Analytical Methods in Chemistry،Vol. 2008, no. 2008, pp.1-6.
https://search.emarefa.net/detail/BIM-987970
نمط استشهاد الجمعية الأمريكية للغات الحديثة (MLA)
Brown, Richard J. C.…[et al.]. On the Optimum Sampling Time for the Measurement of Pollutants in Ambient Air. Journal of Analytical Methods in Chemistry No. 2008 (2008), pp.1-6.
https://search.emarefa.net/detail/BIM-987970
نمط استشهاد الجمعية الطبية الأمريكية (AMA)
Brown, Richard J. C.& Hood, Daniel& Brown, Andrew S.. On the Optimum Sampling Time for the Measurement of Pollutants in Ambient Air. Journal of Analytical Methods in Chemistry. 2008. Vol. 2008, no. 2008, pp.1-6.
https://search.emarefa.net/detail/BIM-987970
نوع البيانات
مقالات
لغة النص
الإنجليزية
الملاحظات
Includes bibliographical references
رقم السجل
BIM-987970
قاعدة معامل التأثير والاستشهادات المرجعية العربي "ارسيف Arcif"
أضخم قاعدة بيانات عربية للاستشهادات المرجعية للمجلات العلمية المحكمة الصادرة في العالم العربي
تقوم هذه الخدمة بالتحقق من التشابه أو الانتحال في الأبحاث والمقالات العلمية والأطروحات الجامعية والكتب والأبحاث باللغة العربية، وتحديد درجة التشابه أو أصالة الأعمال البحثية وحماية ملكيتها الفكرية. تعرف اكثر