Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact TonoPachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure

Joint Authors

Shimizu, Kimiya
Fujimura, Fusako
Kamiya, Kazutaka
Fujiwara, Kazuko
Shoji, Nobuyuki

Source

BioMed Research International

Issue

Vol. 2013, Issue 2013 (31 Dec. 2013), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2013-10-03

Country of Publication

Egypt

No. of Pages

5

Main Subjects

Medicine

Abstract EN

Purpose.

To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP.

Methods.

Forty-six eyes of healthy volunteers were measured by two examiners.

Three consecutive measurements per eye were performed.

Repeatability was assessed using the coefficient of variation, and reproducibility was assessed using Bland-Altman plots.

Linear correlations were used to determine agreement between CCT and noncorrected IOP and CCT and corrected IOP, which was calculated using a formula built into the NT-530P.

Results.

The coefficient of variation for IOP was 6.4% and for CCT was 0.4%.

The 95% limits of agreement between examiners were −0.17±1.42 mmHg (range: −2.95 to 2.61 mmHg) for IOP, −0.93±4.37 μm (range: −9.50 to 7.64 μm) for CCT.

The corrected IOP was significantly higher than the noncorrected IOP (P=0.010.3).

The noncorrected IOP significantly correlated with CCT (r=−0.4883, P=0.0006).

The corrected IOP showed no significant correlation with CCT (r=−0.0285, P=0.8509).

Conclusions.

NT-530P offered repeatability and reproducibility in both IOP and CCT measurements.

The corrected IOP calculated using the NT-530P was independent of the CCT, suggesting that this IOP may be less influenced by the central corneal thickness.

American Psychological Association (APA)

Fujimura, Fusako& Kamiya, Kazutaka& Fujiwara, Kazuko& Shoji, Nobuyuki& Shimizu, Kimiya. 2013. Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact TonoPachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure. BioMed Research International،Vol. 2013, no. 2013, pp.1-5.
https://search.emarefa.net/detail/BIM-1004125

Modern Language Association (MLA)

Fujimura, Fusako…[et al.]. Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact TonoPachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure. BioMed Research International No. 2013 (2013), pp.1-5.
https://search.emarefa.net/detail/BIM-1004125

American Medical Association (AMA)

Fujimura, Fusako& Kamiya, Kazutaka& Fujiwara, Kazuko& Shoji, Nobuyuki& Shimizu, Kimiya. Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact TonoPachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure. BioMed Research International. 2013. Vol. 2013, no. 2013, pp.1-5.
https://search.emarefa.net/detail/BIM-1004125

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1004125