![](/images/graphics-bg.png)
Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
Joint Authors
Chlup, R.
Doubravova, B.
Bartek, J.
Zapletalova, J.
Krystynik, O.
Prochazka, V.
Source
Issue
Vol. 2013, Issue 2013 (31 Dec. 2013), pp.1-11, 11 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2013-12-19
Country of Publication
Egypt
No. of Pages
11
Main Subjects
Abstract EN
Aim of this trial was to assess (1) the accuracy and precision of electrochemistry-based glucometers CONTOURLINK, CALLA, and LINUS and (2) the diabetes control using Ambulatory Glycaemic Profiles (AGP) as markers of therapeutic effectiveness.
Glucometers and COBAS INTEGRA 400 Plus analyzer were used by one laboratory professional to estimate P-glucose (PG) in 112 out-patients.
There were 112 sets of 12 PG estimations analyzed.
In each set, means of 3 capillary PG estimations on 3 respective glucometers and on INTEGRA analyzer were calculated.
The statistical program SPSS, v.
15, was applied.
The mean INTEGRA PG values ranged from 2.7 to 25.3 mmol/L.
There were strong correlations between mean PG on INTEGRA versus CONTOURLINK, versus CALLA, and versus LINUS; PG deviations from INTEGRA were mostly within the range ±15%.
Wilcoxon Signed Rank Test revealed differences between CONTOURLINK-INTEGRA and LINUS-INTEGRA; CALLA-INTEGRA showed no difference.
SD INTEGRA = 0.061 mmol/L, SD CONTOURLINK/SD CALLA/SD LINUS were 0.256/0.290/0.286 mmol/L.
All patients were trained to perform defined 10-point PG profiles to adapt food, exercise, and insulin doses.
The PG differences between all tested glucometers and reference values were in borderline of ISO 15197 but worthy of consideration.
AGP are helpful markers of diabetes control.
American Psychological Association (APA)
Chlup, R.& Doubravova, B.& Bartek, J.& Zapletalova, J.& Krystynik, O.& Prochazka, V.. 2013. Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA). Disease Markers،Vol. 2013, no. 2013, pp.1-11.
https://search.emarefa.net/detail/BIM-1005857
Modern Language Association (MLA)
Chlup, R.…[et al.]. Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA). Disease Markers No. 2013 (2013), pp.1-11.
https://search.emarefa.net/detail/BIM-1005857
American Medical Association (AMA)
Chlup, R.& Doubravova, B.& Bartek, J.& Zapletalova, J.& Krystynik, O.& Prochazka, V.. Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA). Disease Markers. 2013. Vol. 2013, no. 2013, pp.1-11.
https://search.emarefa.net/detail/BIM-1005857
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-1005857