Abstract EN
A buddy wire is often used to aid in the delivery of balloons and stents when negotiating tortuous or calcified vessels.
We present a planned two-stent mini-crush intervention complicated by entanglement of the buddy wire with the second stent and subsequent distortion of the stent within the guiding catheter.
Based on this case, we suggest removing the buddy wire immediately after successful positioning of the first stent, because entrapment with a second stent is possible and may lead to challenging situations in a simultaneous two-stent strategy.
American Psychological Association (APA)
Vorpahl, Marc& Seyfarth, Melchior& Tiroch, Klaus. 2014. Entanglement due to Delayed Removal of a Buddy Wire. Case Reports in Cardiology،Vol. 2014, no. 2014, pp.1-3.
https://search.emarefa.net/detail/BIM-1016869
Modern Language Association (MLA)
Vorpahl, Marc…[et al.]. Entanglement due to Delayed Removal of a Buddy Wire. Case Reports in Cardiology No. 2014 (2014), pp.1-3.
https://search.emarefa.net/detail/BIM-1016869
American Medical Association (AMA)
Vorpahl, Marc& Seyfarth, Melchior& Tiroch, Klaus. Entanglement due to Delayed Removal of a Buddy Wire. Case Reports in Cardiology. 2014. Vol. 2014, no. 2014, pp.1-3.
https://search.emarefa.net/detail/BIM-1016869
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-1016869