Assessment of Patients with a DePuy ASR Metal-on-Metal Hip Replacement: Results of Applying the Guidelines of the Spanish Society of Hip Surgery in a Tertiary Referral Hospital

Joint Authors

García, Sebastián
Bori, Guillem
Riba, Josep
Combalía, A.
Fernández-Valencia, J. A.
Gallart, Xavier

Source

Advances in Orthopedics

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-4, 4 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-11-09

Country of Publication

Egypt

No. of Pages

4

Main Subjects

Medicine

Abstract EN

The prognosis associated with the DePuy ASR hip cup is poor and varies according to the series.

This implant was withdrawn from use in 2010 and all patients needed to be assessed.

We present the results of the assessment of our patients treated with this device, according to the Spanish Society of Hip Surgery (SECCA) algorithm published in 2011.

This retrospective study evaluates 83 consecutive ASR cups, followed up at a mean of 2.9 years.

Serum levels of chromium and cobalt, as well as the acetabular abduction angle, were determined in order to assess their possible correlation with failure, defined as the need for revision surgery.

The mean Harris Hip Score was 83.2 (range 42–97).

Eight arthroplasties (13.3%) required revision due to persistent pain and/or elevated serum levels of chromium/cobalt.

All the cups had a correct abduction angle, and there was no correlation between elevated serum levels of metal ions and implant failure.

Since two previous ASR implants were exchanged previously to the recall, the revision rate for ASR cups in our centre is 18.2% at 2.9 years.

American Psychological Association (APA)

Fernández-Valencia, J. A.& Gallart, Xavier& Bori, Guillem& García, Sebastián& Combalía, A.& Riba, Josep. 2014. Assessment of Patients with a DePuy ASR Metal-on-Metal Hip Replacement: Results of Applying the Guidelines of the Spanish Society of Hip Surgery in a Tertiary Referral Hospital. Advances in Orthopedics،Vol. 2014, no. 2014, pp.1-4.
https://search.emarefa.net/detail/BIM-1034350

Modern Language Association (MLA)

Fernández-Valencia, J. A.…[et al.]. Assessment of Patients with a DePuy ASR Metal-on-Metal Hip Replacement: Results of Applying the Guidelines of the Spanish Society of Hip Surgery in a Tertiary Referral Hospital. Advances in Orthopedics No. 2014 (2014), pp.1-4.
https://search.emarefa.net/detail/BIM-1034350

American Medical Association (AMA)

Fernández-Valencia, J. A.& Gallart, Xavier& Bori, Guillem& García, Sebastián& Combalía, A.& Riba, Josep. Assessment of Patients with a DePuy ASR Metal-on-Metal Hip Replacement: Results of Applying the Guidelines of the Spanish Society of Hip Surgery in a Tertiary Referral Hospital. Advances in Orthopedics. 2014. Vol. 2014, no. 2014, pp.1-4.
https://search.emarefa.net/detail/BIM-1034350

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1034350