Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay

Joint Authors

Freeman, James
Wilson, Kimberly
Spears, Ryan
Shalhoub, Victoria
Sibley, Paul

Source

International Journal of Endocrinology

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-12, 12 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-06-19

Country of Publication

Egypt

No. of Pages

12

Main Subjects

Biology

Abstract EN

Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D3 and 25(OH)D2].

This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent reference method procedure) and a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for various populations with different levels of vitamin D binding protein (DBP).

Total serum 25(OH)D concentrations were measured for 36 pregnant women, 40 hemodialysis patients, and 30 samples (DBP-spiked or not) from healthy subjects.

ELISA measured DBP levels.

The mean serum DBP concentrations were higher for pregnancy (415 μg/mL) and lower for hemodialysis subjects (198 μg/mL) than for healthy subjects and were highest for spiked serum (545 μg/mL).

The average bias between the ADVIA Centaur assay and the LC-MS/MS method was −1.4% (healthy), −6.1% (pregnancy), and 4.4% (hemodialysis).

The slightly greater bias for samples from some pregnancy and hemodialysis subjects with serum DBP levels outside of the normal healthy range fell within a clinically acceptable range—reflected by analysis of their low-range (≤136 μg/mL), medium-range (137–559 μg/mL), and high-range (≥560 μg/mL) DBP groups.

Thus, the ADVIA Centaur Vitamin D Total assay demonstrates acceptable performance compared with an LC-MS/MS method for populations containing different amounts of DBP.

American Psychological Association (APA)

Freeman, James& Wilson, Kimberly& Spears, Ryan& Shalhoub, Victoria& Sibley, Paul. 2014. Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay. International Journal of Endocrinology،Vol. 2014, no. 2014, pp.1-12.
https://search.emarefa.net/detail/BIM-1036564

Modern Language Association (MLA)

Freeman, James…[et al.]. Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay. International Journal of Endocrinology No. 2014 (2014), pp.1-12.
https://search.emarefa.net/detail/BIM-1036564

American Medical Association (AMA)

Freeman, James& Wilson, Kimberly& Spears, Ryan& Shalhoub, Victoria& Sibley, Paul. Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay. International Journal of Endocrinology. 2014. Vol. 2014, no. 2014, pp.1-12.
https://search.emarefa.net/detail/BIM-1036564

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1036564