Correction for Eddy Current-Induced Echo-Shifting Effect in Partial-Fourier Diffusion Tensor Imaging

Joint Authors

Chen, Nan-Kuei
Truong, Trong-Kha
Song, Allen W.

Source

BioMed Research International

Issue

Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-12, 12 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2015-08-30

Country of Publication

Egypt

No. of Pages

12

Main Subjects

Medicine

Abstract EN

In most diffusion tensor imaging (DTI) studies, images are acquired with either a partial-Fourier or a parallel partial-Fourier echo-planar imaging (EPI) sequence, in order to shorten the echo time and increase the signal-to-noise ratio (SNR).

However, eddy currents induced by the diffusion-sensitizing gradients can often lead to a shift of the echo in k-space, resulting in three distinct types of artifacts in partial-Fourier DTI.

Here, we present an improved DTI acquisition and reconstruction scheme, capable of generating high-quality and high-SNR DTI data without eddy current-induced artifacts.

This new scheme consists of three components, respectively, addressing the three distinct types of artifacts.

First, a k-space energy-anchored DTI sequence is designed to recover eddy current-induced signal loss (i.e., Type 1 artifact).

Second, a multischeme partial-Fourier reconstruction is used to eliminate artificial signal elevation (i.e., Type 2 artifact) associated with the conventional partial-Fourier reconstruction.

Third, a signal intensity correction is applied to remove artificial signal modulations due to eddy current-induced erroneous T2*-weighting (i.e., Type 3 artifact).

These systematic improvements will greatly increase the consistency and accuracy of DTI measurements, expanding the utility of DTI in translational applications where quantitative robustness is much needed.

American Psychological Association (APA)

Truong, Trong-Kha& Song, Allen W.& Chen, Nan-Kuei. 2015. Correction for Eddy Current-Induced Echo-Shifting Effect in Partial-Fourier Diffusion Tensor Imaging. BioMed Research International،Vol. 2015, no. 2015, pp.1-12.
https://search.emarefa.net/detail/BIM-1054538

Modern Language Association (MLA)

Truong, Trong-Kha…[et al.]. Correction for Eddy Current-Induced Echo-Shifting Effect in Partial-Fourier Diffusion Tensor Imaging. BioMed Research International No. 2015 (2015), pp.1-12.
https://search.emarefa.net/detail/BIM-1054538

American Medical Association (AMA)

Truong, Trong-Kha& Song, Allen W.& Chen, Nan-Kuei. Correction for Eddy Current-Induced Echo-Shifting Effect in Partial-Fourier Diffusion Tensor Imaging. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-12.
https://search.emarefa.net/detail/BIM-1054538

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1054538