A Novel Simple Phantom for Verifying the Dose of Radiation Therapy

Joint Authors

Hsu, S. M.
Shiau, An-Cheng
Lee, J. H.
Chang, L. T.
Chen, C. W.
Li, W. J.
Lee, M. S.
Liao, Y. J.

Source

BioMed Research International

Issue

Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2015-03-26

Country of Publication

Egypt

No. of Pages

5

Main Subjects

Medicine

Abstract EN

A standard protocol of dosimetric measurements is used by the organizations responsible for verifying that the doses delivered in radiation-therapy institutions are within authorized limits.

This study evaluated a self-designed simple auditing phantom for use in verifying the dose of radiation therapy; the phantom design, dose audit system, and clinical tests are described.

Thermoluminescent dosimeters (TLDs) were used as postal dosimeters, and mailable phantoms were produced for use in postal audits.

Correction factors are important for converting TLD readout values from phantoms into the absorbed dose in water.

The phantom scatter correction factor was used to quantify the difference in the scattered dose between a solid water phantom and homemade phantoms; its value ranged from 1.084 to 1.031.

The energy-dependence correction factor was used to compare the TLD readout of the unit dose irradiated by audit beam energies with 60Co in the solid water phantom; its value was 0.99 to 1.01.

The setup-condition factor was used to correct for differences in dose-output calibration conditions.

Clinical tests of the device calibrating the dose output revealed that the dose deviation was within 3%.

Therefore, our homemade phantoms and dosimetric system can be applied for accurately verifying the doses applied in radiation-therapy institutions.

American Psychological Association (APA)

Lee, J. H.& Chang, L. T.& Shiau, An-Cheng& Chen, C. W.& Liao, Y. J.& Li, W. J.…[et al.]. 2015. A Novel Simple Phantom for Verifying the Dose of Radiation Therapy. BioMed Research International،Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1057298

Modern Language Association (MLA)

Lee, J. H.…[et al.]. A Novel Simple Phantom for Verifying the Dose of Radiation Therapy. BioMed Research International No. 2015 (2015), pp.1-5.
https://search.emarefa.net/detail/BIM-1057298

American Medical Association (AMA)

Lee, J. H.& Chang, L. T.& Shiau, An-Cheng& Chen, C. W.& Liao, Y. J.& Li, W. J.…[et al.]. A Novel Simple Phantom for Verifying the Dose of Radiation Therapy. BioMed Research International. 2015. Vol. 2015, no. 2015, pp.1-5.
https://search.emarefa.net/detail/BIM-1057298

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1057298