Adaptive Autoregressive Model for Reduction of Noise in SPECT

Joint Authors

Takalo, Reijo
Hytti, Heli
Ihalainen, Heimo
Sohlberg, Antti

Source

Computational and Mathematical Methods in Medicine

Issue

Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-9, 9 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2015-05-18

Country of Publication

Egypt

No. of Pages

9

Main Subjects

Medicine

Abstract EN

This paper presents improved autoregressive modelling (AR) to reduce noise in SPECT images.

An AR filter was applied to prefilter projection images and postfilter ordered subset expectation maximisation (OSEM) reconstruction images (AR-OSEM-AR method).

The performance of this method was compared with filtered back projection (FBP) preceded by Butterworth filtering (BW-FBP method) and the OSEM reconstruction method followed by Butterworth filtering (OSEM-BW method).

A mathematical cylinder phantom was used for the study.

It consisted of hot and cold objects.

The tests were performed using three simulated SPECT datasets.

Image quality was assessed by means of the percentage contrast resolution (CR%) and the full width at half maximum (FWHM) of the line spread functions of the cylinders.

The BW-FBP method showed the highest CR% values and the AR-OSEM-AR method gave the lowest CR% values for cold stacks.

In the analysis of hot stacks, the BW-FBP method had higher CR% values than the OSEM-BW method.

The BW-FBP method exhibited the lowest FWHM values for cold stacks and the AR-OSEM-AR method for hot stacks.

In conclusion, the AR-OSEM-AR method is a feasible way to remove noise from SPECT images.

It has good spatial resolution for hot objects.

American Psychological Association (APA)

Takalo, Reijo& Hytti, Heli& Ihalainen, Heimo& Sohlberg, Antti. 2015. Adaptive Autoregressive Model for Reduction of Noise in SPECT. Computational and Mathematical Methods in Medicine،Vol. 2015, no. 2015, pp.1-9.
https://search.emarefa.net/detail/BIM-1057918

Modern Language Association (MLA)

Takalo, Reijo…[et al.]. Adaptive Autoregressive Model for Reduction of Noise in SPECT. Computational and Mathematical Methods in Medicine No. 2015 (2015), pp.1-9.
https://search.emarefa.net/detail/BIM-1057918

American Medical Association (AMA)

Takalo, Reijo& Hytti, Heli& Ihalainen, Heimo& Sohlberg, Antti. Adaptive Autoregressive Model for Reduction of Noise in SPECT. Computational and Mathematical Methods in Medicine. 2015. Vol. 2015, no. 2015, pp.1-9.
https://search.emarefa.net/detail/BIM-1057918

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID