Abstract EN

Modularity of the components in total hip arthroplasty has had an increase in popularity in the last decades.

We present the case of a 53-year-old man with a history of avascular necrosis of the femoral head due to a hypophyseal adenoma.

A total hip modular arthroplasty was implanted.

Three and a half years after the surgery the patient attended the emergency room due to acute left hip pain with no prior traumatism.

Radiological examination confirmed a fracture of the modular neck.

A revision surgery was performed finding an important pseudotumoral well-organized periprosthetic tissue reaction.

Through an extended trochanteric osteotomy the femoral component was removed, and a straight-stem revision prosthesis implanted.

There are several potential advantages when using modularity in total hip arthroplasty that surgeons may benefit from, but complications have arisen and must be addressed.

Various circumstances such as large femoral head with a long varus neck, corrosion, patient’s BMI, and activity level may participate in creating the necessary environment for fatigue failure of the implant.

American Psychological Association (APA)

Hernandez, A.& Gargallo-Margarit, Albert& Barro, Victor& Gallardo-Calero, I.& Sallent, A.. 2015. Fracture of the Modular Neck in Total Hip Arthroplasty. Case Reports in Orthopedics،Vol. 2015, no. 2015, pp.1-4.
https://search.emarefa.net/detail/BIM-1059507

Modern Language Association (MLA)

Hernandez, A.…[et al.]. Fracture of the Modular Neck in Total Hip Arthroplasty. Case Reports in Orthopedics No. 2015 (2015), pp.1-4.
https://search.emarefa.net/detail/BIM-1059507

American Medical Association (AMA)

Hernandez, A.& Gargallo-Margarit, Albert& Barro, Victor& Gallardo-Calero, I.& Sallent, A.. Fracture of the Modular Neck in Total Hip Arthroplasty. Case Reports in Orthopedics. 2015. Vol. 2015, no. 2015, pp.1-4.
https://search.emarefa.net/detail/BIM-1059507

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1059507