Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry
Joint Authors
Huang, Jinhai
Wang, Qinmei
Wang, Chengfang
Lu, Weicong
Gao, Rongrong
Li, Yuanguang
Zhao, Yune
Savini, G.
Source
Issue
Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-6, 6 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2015-02-24
Country of Publication
Egypt
No. of Pages
6
Main Subjects
Abstract EN
Purpose.
To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry.
Methods.
Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken.
The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility.
Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA).
Results.
The intraoperators TRT and CoV were <19 μm and 2.0%, respectively.
The interoperators TRT and CoV were <12 μm and 1.0%, respectively, and ICC was >0.90.
The mean CCT and TCT measurements using the TMS-5 were 15.97 μm (95% LoA from −26.42 to −5.52 μm) and 20.32 μm (95% LoA from −30.67 to −9.97 μm) smaller, respectively, than those using US pachymetry.
Conclusions.
The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results.
Caution is warranted before using these techniques interchangeably.
American Psychological Association (APA)
Huang, Jinhai& Savini, G.& Wang, Chengfang& Lu, Weicong& Gao, Rongrong& Li, Yuanguang…[et al.]. 2015. Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry. Journal of Ophthalmology،Vol. 2015, no. 2015, pp.1-6.
https://search.emarefa.net/detail/BIM-1069701
Modern Language Association (MLA)
Huang, Jinhai…[et al.]. Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry. Journal of Ophthalmology No. 2015 (2015), pp.1-6.
https://search.emarefa.net/detail/BIM-1069701
American Medical Association (AMA)
Huang, Jinhai& Savini, G.& Wang, Chengfang& Lu, Weicong& Gao, Rongrong& Li, Yuanguang…[et al.]. Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry. Journal of Ophthalmology. 2015. Vol. 2015, no. 2015, pp.1-6.
https://search.emarefa.net/detail/BIM-1069701
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-1069701