Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry

Joint Authors

Huang, Jinhai
Wang, Qinmei
Wang, Chengfang
Lu, Weicong
Gao, Rongrong
Li, Yuanguang
Zhao, Yune
Savini, G.

Source

Journal of Ophthalmology

Issue

Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-6, 6 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2015-02-24

Country of Publication

Egypt

No. of Pages

6

Main Subjects

Medicine

Abstract EN

Purpose.

To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry.

Methods.

Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken.

The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility.

Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA).

Results.

The intraoperators TRT and CoV were <19 μm and 2.0%, respectively.

The interoperators TRT and CoV were <12 μm and 1.0%, respectively, and ICC was >0.90.

The mean CCT and TCT measurements using the TMS-5 were 15.97 μm (95% LoA from −26.42 to −5.52 μm) and 20.32 μm (95% LoA from −30.67 to −9.97 μm) smaller, respectively, than those using US pachymetry.

Conclusions.

The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results.

Caution is warranted before using these techniques interchangeably.

American Psychological Association (APA)

Huang, Jinhai& Savini, G.& Wang, Chengfang& Lu, Weicong& Gao, Rongrong& Li, Yuanguang…[et al.]. 2015. Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry. Journal of Ophthalmology،Vol. 2015, no. 2015, pp.1-6.
https://search.emarefa.net/detail/BIM-1069701

Modern Language Association (MLA)

Huang, Jinhai…[et al.]. Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry. Journal of Ophthalmology No. 2015 (2015), pp.1-6.
https://search.emarefa.net/detail/BIM-1069701

American Medical Association (AMA)

Huang, Jinhai& Savini, G.& Wang, Chengfang& Lu, Weicong& Gao, Rongrong& Li, Yuanguang…[et al.]. Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry. Journal of Ophthalmology. 2015. Vol. 2015, no. 2015, pp.1-6.
https://search.emarefa.net/detail/BIM-1069701

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1069701