Multivariate Analysis of the Ocular Response Analyzer’s Corneal Deformation Response Curve for Early Keratoconus Detection

Joint Authors

Galletti, Jonatán D.
Ruiseñor Vázquez, Pablo R.
Fuentes Bonthoux, Fernando
Pförtner, Tomás
Galletti, Jeremías G.

Source

Journal of Ophthalmology

Issue

Vol. 2015, Issue 2015 (31 Dec. 2015), pp.1-8, 8 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2015-05-14

Country of Publication

Egypt

No. of Pages

8

Main Subjects

Medicine

Abstract EN

Purpose.

To thoroughly analyze corneal deformation responses curves obtained by Ocular Response Analyzer (ORA) testing in order to improve subclinical keratoconus detection.

Methods.

Observational case series of 87 control and 73 subclinical keratoconus eyes.

Examination included corneal topography, tomography, and biomechanical testing with ORA.

Factor analysis, logistic regression, and receiver operating characteristic curves were used to extract combinations of 45 corneal waveform descriptors.

Main outcome measures were corneal-thickness-corrected corneal resistance factor (ccCRF), combinations of corneal descriptors, and their diagnostic performance.

Results.

Thirty-seven descriptors differed significantly in means between groups, and among them ccCRF afforded the highest individual diagnostic performance.

Factor analysis identified first- and second-peak related descriptors as the most variable one.

However, conventional biomechanical descriptors corneal resistance factor and hysteresis differed the most between control and keratoconic eyes.

A combination of three factors including several corneal descriptors did not show better diagnostic performance than a combination of conventional indices.

Conclusion.

Multivariate analysis of ORA signals did not surpass simpler models in subclinical keratoconus detection, and there is considerable overlap between normal and ectatic eyes irrespective of the analysis model.

Conventional biomechanical indices seem to already provide the best performance when appropriately considered.

American Psychological Association (APA)

Galletti, Jonatán D.& Ruiseñor Vázquez, Pablo R.& Fuentes Bonthoux, Fernando& Pförtner, Tomás& Galletti, Jeremías G.. 2015. Multivariate Analysis of the Ocular Response Analyzer’s Corneal Deformation Response Curve for Early Keratoconus Detection. Journal of Ophthalmology،Vol. 2015, no. 2015, pp.1-8.
https://search.emarefa.net/detail/BIM-1069755

Modern Language Association (MLA)

Galletti, Jonatán D.…[et al.]. Multivariate Analysis of the Ocular Response Analyzer’s Corneal Deformation Response Curve for Early Keratoconus Detection. Journal of Ophthalmology No. 2015 (2015), pp.1-8.
https://search.emarefa.net/detail/BIM-1069755

American Medical Association (AMA)

Galletti, Jonatán D.& Ruiseñor Vázquez, Pablo R.& Fuentes Bonthoux, Fernando& Pförtner, Tomás& Galletti, Jeremías G.. Multivariate Analysis of the Ocular Response Analyzer’s Corneal Deformation Response Curve for Early Keratoconus Detection. Journal of Ophthalmology. 2015. Vol. 2015, no. 2015, pp.1-8.
https://search.emarefa.net/detail/BIM-1069755

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1069755