Two-Year Follow-Up after Contact Force Sensing Radiofrequency Catheter and Second-Generation Cryoballoon Ablation for Paroxysmal Atrial Fibrillation: A Comparative Single Centre Study

Joint Authors

Kardos, Attila
Kis, Zsuzsanna
Som, Zoltan
Nagy, Zsofia
Foldesi, Csaba

Source

BioMed Research International

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-6, 6 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-05-22

Country of Publication

Egypt

No. of Pages

6

Main Subjects

Medicine

Abstract EN

Background.

There are little comparative data on catheter ablation of paroxysmal atrial fibrillation (AF) using the contact force radiofrequency (CF-RF) catheter versus the second-generation cryoballoon (CB2).

Methods and results.

This is a single center, retrospective, nonrandomized study of 98 patients with symptomatic, drug-refractory paroxysmal AF who underwent their first PVI ablation using either the CB2 (n=40) or CF-RF (n=58).

The mean age was 60 years with 63% men, a mean LA size of 42 mm.

The procedure duration (74±17 versus 120±49 minutes p<0.05) was shorter for CB2 group; the fluoroscopy time (14±17 versus 16±5 minutes, p=0.45) was similar.

Complete PVI was achieved in 96% of patients with RF-CF and 98% with CB2.

Phrenic nerve palsies (2 transient and 1 persistent) occurred exclusively in the CB2 group and 1 severe, nonlethal complication (pericardial tamponade) occurred in the CF-RF group.

At 24-month follow-up, the success rate, defined as freedom from AF/atrial tachycardia (AT) after a single procedure without antiarrhythmic drug, was comparable in CF-RF group and CB2 group (65.5% versus 67%, resp., log rank p=0.54).

Conclusion.

Both the CB2 and the RF-CF ablation appeared safe; the success rate at 2 years was comparable between both technologies.

American Psychological Association (APA)

Kardos, Attila& Kis, Zsuzsanna& Som, Zoltan& Nagy, Zsofia& Foldesi, Csaba. 2016. Two-Year Follow-Up after Contact Force Sensing Radiofrequency Catheter and Second-Generation Cryoballoon Ablation for Paroxysmal Atrial Fibrillation: A Comparative Single Centre Study. BioMed Research International،Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098499

Modern Language Association (MLA)

Kardos, Attila…[et al.]. Two-Year Follow-Up after Contact Force Sensing Radiofrequency Catheter and Second-Generation Cryoballoon Ablation for Paroxysmal Atrial Fibrillation: A Comparative Single Centre Study. BioMed Research International No. 2016 (2016), pp.1-6.
https://search.emarefa.net/detail/BIM-1098499

American Medical Association (AMA)

Kardos, Attila& Kis, Zsuzsanna& Som, Zoltan& Nagy, Zsofia& Foldesi, Csaba. Two-Year Follow-Up after Contact Force Sensing Radiofrequency Catheter and Second-Generation Cryoballoon Ablation for Paroxysmal Atrial Fibrillation: A Comparative Single Centre Study. BioMed Research International. 2016. Vol. 2016, no. 2016, pp.1-6.
https://search.emarefa.net/detail/BIM-1098499

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1098499