One-Piece Zirconia Ceramic versus Titanium Implants in the Jaw and Femur of a Sheep Model: A Pilot Study

Joint Authors

Duncan, Warwick J.
Siddiqi, A.
De Silva, R. K.
Zafar, S.

Source

BioMed Research International

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-10, 10 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-12-12

Country of Publication

Egypt

No. of Pages

10

Main Subjects

Medicine

Abstract EN

Reports have documented titanium (Ti) hypersensitivity after dental implant treatment.

Alternative materials have been suggested including zirconia (Zr) ceramics, which have shown predictable osseointegration in animal studies and appear free of immune responses.

The aim of the research was to investigate the bone-to-implant contact (BIC) of one-piece Zr, compared with one-piece Ti implants, placed in the jaws and femurs of domestic sheep.

Ten New Zealand mixed breed sheep were used.

A One-piece prototype Ti (control) and one Zr (test) implant were placed in the mandible, and one of each implant (Ti and Zr) was placed into the femoral epicondyle of each animal.

The femur implants were submerged and unloaded; the mandibular implants were placed using a one-stage transgingival protocol and were nonsubmerged.

After a healing period of 12 weeks, %BIC was measured.

The overall survival rate for mandibular and femur implants combined was 87.5%.

%BIC was higher for Zr implants versus Ti implants in the femur (85.5%, versus 78.9%) (p=0.002).

Zirconia implants in the mandible showed comparable %BIC to titanium implants (72.2%, versus 60.3%) (p=0.087).

High failure rate of both Zr and Ti one-piece implants in the jaw could be attributed to the one-piece design and surface characteristics of the implant that could have influenced osseointegration.

Further clinical trials are recommended to evaluate the performance of zirconia implants under loading conditions.

American Psychological Association (APA)

Siddiqi, A.& Duncan, Warwick J.& De Silva, R. K.& Zafar, S.. 2016. One-Piece Zirconia Ceramic versus Titanium Implants in the Jaw and Femur of a Sheep Model: A Pilot Study. BioMed Research International،Vol. 2016, no. 2016, pp.1-10.
https://search.emarefa.net/detail/BIM-1098555

Modern Language Association (MLA)

Siddiqi, A.…[et al.]. One-Piece Zirconia Ceramic versus Titanium Implants in the Jaw and Femur of a Sheep Model: A Pilot Study. BioMed Research International No. 2016 (2016), pp.1-10.
https://search.emarefa.net/detail/BIM-1098555

American Medical Association (AMA)

Siddiqi, A.& Duncan, Warwick J.& De Silva, R. K.& Zafar, S.. One-Piece Zirconia Ceramic versus Titanium Implants in the Jaw and Femur of a Sheep Model: A Pilot Study. BioMed Research International. 2016. Vol. 2016, no. 2016, pp.1-10.
https://search.emarefa.net/detail/BIM-1098555

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1098555