Contrast-Based 3D2D Registration of the Left Atrium: Fast versus Consistent

Joint Authors

Hoffmann, Matthias
Kowalewski, Christopher
Kurzidim, Klaus
Strobel, Norbert
Maier, Andreas
Hornegger, Joachim

Source

International Journal of Biomedical Imaging

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-15, 15 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-03-08

Country of Publication

Egypt

No. of Pages

15

Main Subjects

Medicine

Abstract EN

For augmented fluoroscopy during cardiac ablation, a preoperatively acquired 3D model of a patient’s left atrium (LA) can be registered to X-ray images recorded during a contrast agent (CA) injection.

An automatic registration method that works also for small amounts of CA is desired.

We propose two similarity measures: The first focuses on edges of the patient anatomy.

The second computes a contrast agent distribution estimate (CADE) inside the 3D model and rates its consistency with the CA as seen in biplane fluoroscopic images.

Moreover, temporal filtering on the obtained registration results of a sequence is applied using a Markov chain framework.

Evaluation was performed on 11 well-contrasted clinical angiographic sequences and 10 additional sequences with less CA.

For well-contrasted sequences, the error for all 73 frames was 7.9 ± 6.3 mm and it dropped to 4.6 ± 4.0 mm when registering to an automatically selected, well enhanced frame in each sequence.

Temporal filtering reduced the error for all frames from 7.9 ± 6.3 mm to 5.7 ± 4.6 mm.

The error was typically higher if less CA was used.

A combination of both similarity measures outperforms a previously proposed similarity measure.

The mean accuracy for well contrasted sequences is in the range of other proposed manual registration methods.

American Psychological Association (APA)

Hoffmann, Matthias & Kowalewski, Christopher& Maier, Andreas& Kurzidim, Klaus& Strobel, Norbert& Hornegger, Joachim. 2016. Contrast-Based 3D2D Registration of the Left Atrium: Fast versus Consistent. International Journal of Biomedical Imaging،Vol. 2016, no. 2016, pp.1-15.
https://search.emarefa.net/detail/BIM-1105436

Modern Language Association (MLA)

Hoffmann, Matthias…[et al.]. Contrast-Based 3D2D Registration of the Left Atrium: Fast versus Consistent. International Journal of Biomedical Imaging No. 2016 (2016), pp.1-15.
https://search.emarefa.net/detail/BIM-1105436

American Medical Association (AMA)

Hoffmann, Matthias & Kowalewski, Christopher& Maier, Andreas& Kurzidim, Klaus& Strobel, Norbert& Hornegger, Joachim. Contrast-Based 3D2D Registration of the Left Atrium: Fast versus Consistent. International Journal of Biomedical Imaging. 2016. Vol. 2016, no. 2016, pp.1-15.
https://search.emarefa.net/detail/BIM-1105436

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1105436