Catalase-Based Modified Graphite Electrode for Hydrogen Peroxide Detection in Different Beverages

Joint Authors

Antiochia, Riccarda
Bollella, Paolo
Favero, Gabriele
Mazzei, Franco
Fusco, Giovanni
Tortolini, Cristina

Source

Journal of Analytical Methods in Chemistry

Issue

Vol. 2016, Issue 2016 (31 Dec. 2016), pp.1-12, 12 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2016-12-18

Country of Publication

Egypt

No. of Pages

12

Main Subjects

Chemistry

Abstract EN

A catalase-based (NAF/MWCNTs) nanocomposite film modified glassy carbon electrode for hydrogen peroxide (H2O2) detection was developed.

The developed biosensor was characterized in terms of its bioelectrochemical properties.

Cyclic voltammetry (CV) technique was employed to study the redox features of the enzyme in the absence and in the presence of nanomaterials dispersed in Nafion® polymeric solution.

The electron transfer coefficient, α, and the electron transfer rate constant, ks, were found to be 0.42 and 1.71 s−1, at pH 7.0, respectively.

Subsequently, the same modification steps were applied to mesoporous graphite screen-printed electrodes.

Also, these electrodes were characterized in terms of their main electrochemical and kinetic parameters.

The biosensor performances improved considerably after modification with nanomaterials.

Moreover, the association of Nafion with carbon nanotubes retained the biological activity of the redox protein.

The enzyme electrode response was linear in the range 2.5–1150 μmol L−1, with LOD of 0.83 μmol L−1.

From the experimental data, we can assess the possibility of using the modified biosensor as a useful tool for H2O2 determination in packaged beverages.

American Psychological Association (APA)

Fusco, Giovanni& Bollella, Paolo& Mazzei, Franco& Favero, Gabriele& Antiochia, Riccarda& Tortolini, Cristina. 2016. Catalase-Based Modified Graphite Electrode for Hydrogen Peroxide Detection in Different Beverages. Journal of Analytical Methods in Chemistry،Vol. 2016, no. 2016, pp.1-12.
https://search.emarefa.net/detail/BIM-1107375

Modern Language Association (MLA)

Fusco, Giovanni…[et al.]. Catalase-Based Modified Graphite Electrode for Hydrogen Peroxide Detection in Different Beverages. Journal of Analytical Methods in Chemistry No. 2016 (2016), pp.1-12.
https://search.emarefa.net/detail/BIM-1107375

American Medical Association (AMA)

Fusco, Giovanni& Bollella, Paolo& Mazzei, Franco& Favero, Gabriele& Antiochia, Riccarda& Tortolini, Cristina. Catalase-Based Modified Graphite Electrode for Hydrogen Peroxide Detection in Different Beverages. Journal of Analytical Methods in Chemistry. 2016. Vol. 2016, no. 2016, pp.1-12.
https://search.emarefa.net/detail/BIM-1107375

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1107375