Custom-Made Direct Metal Laser Sintering Titanium Subperiosteal Implants: A Retrospective Clinical Study on 70 Patients

Joint Authors

Dolcini, Giorgio Andrea
Cerea, Mauro

Source

BioMed Research International

Issue

Vol. 2018, Issue 2018 (31 Dec. 2018), pp.1-11, 11 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2018-05-28

Country of Publication

Egypt

No. of Pages

11

Main Subjects

Medicine

Abstract EN

Purpose.

To present a digital technique for the fabrication of custom-made subperiosteal implants and to report on the survival and complication rates encountered when using these fixtures.

Methods.

The data used for this retrospective clinical study were derived from the medical records of five different private dental practices.

Inclusion criteria were patients over the age of 60, treated with custom-made direct metal laser sintering (DMLS) titanium subperiosteal implants (Eagle-Grid®, BTK, Dueville, Vicenza) during a two-year period (2014-2015) and restored with fixed restorations; all enrolled patients needed to have complete pre- and postoperative clinical and radiographic documentation, with at least 2 years of follow-up.

Exclusion criteria were smoking and bruxism.

The main outcomes looked at were implant survival and complications.

Results.

Seventy patients (39 males and 31 females, aged 62-79 years) who had been treated with custom-made DMLS titanium subperiosteal implants were enrolled in this study.

After 2 years of follow-up, three implants were lost due to recurrent, untreatable infections; the survival rate was therefore 95.8% (67/70 implants).

Four patients reported pain/discomfort/swelling after implant placement; the incidence of immediate postoperative complications was therefore 5.7% (4/70 implants).

During the follow-up period, one patient suffered from recurrent infections classified as a biologic complication; the incidence of biologic complications was therefore 1.4% (1/67 surviving implants).

Finally, four patients experienced prosthetic problems with their implant-supported restorations during the provisional phase (fracture of the acrylic restoration) and two patients had ceramic chipping of the definitive restoration; the incidence of prosthetic complications was therefore 8.9% (6/67 surviving implants).

Conclusions.

Within the limits of the present study (limited follow-up time and low number of patients treated, retrospective design), the application of custom-made DMLS titanium subperiosteal implants showed satisfactory implant survival (95.8%) and low complication rates.

Further studies are needed to confirm the positive outcomes found in this research.

American Psychological Association (APA)

Cerea, Mauro& Dolcini, Giorgio Andrea. 2018. Custom-Made Direct Metal Laser Sintering Titanium Subperiosteal Implants: A Retrospective Clinical Study on 70 Patients. BioMed Research International،Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1127276

Modern Language Association (MLA)

Cerea, Mauro& Dolcini, Giorgio Andrea. Custom-Made Direct Metal Laser Sintering Titanium Subperiosteal Implants: A Retrospective Clinical Study on 70 Patients. BioMed Research International No. 2018 (2018), pp.1-11.
https://search.emarefa.net/detail/BIM-1127276

American Medical Association (AMA)

Cerea, Mauro& Dolcini, Giorgio Andrea. Custom-Made Direct Metal Laser Sintering Titanium Subperiosteal Implants: A Retrospective Clinical Study on 70 Patients. BioMed Research International. 2018. Vol. 2018, no. 2018, pp.1-11.
https://search.emarefa.net/detail/BIM-1127276

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1127276