Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty

Joint Authors

Sairyo, Koichi
Mineta, Kazuaki
Okada, Masahiko
Matsumoto, Soshi
Goto, Tomohiro
Hamada, Daisuke

Source

Case Reports in Orthopedics

Issue

Vol. 2017, Issue 2017 (31 Dec. 2017), pp.1-7, 7 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2017-05-16

Country of Publication

Egypt

No. of Pages

7

Main Subjects

Medicine

Abstract EN

We report a rare case of tibial baseplate fracture of Miller-Galante II (MG II) prosthesis.

We examine the factors that may cause such late fracture and review the literature on radiographic analysis and retrieval studies.

A 76-year-old woman, who had undergone bilateral MG II total knee arthroplasty due to rheumatoid arthritis 16 years earlier, presented to our department with a 3-month history of left knee pain.

Plain radiographs revealed severe distortion of the medial tibial component.

During revision knee arthroplasty, we observed severe metallosis in the knee joint, polyethylene insert deformation, and posteromedial coronal baseplate fracture.

After removal of the fractured tray, a bone deficit due to osteolysis was noted.

The revision prosthesis (LCCK, Zimmer-Biomet) was implanted uneventfully.

Four months after revision surgery, the patient was ambulating and had no complications.

The implants on the right side had survived without complications for 17 years.

We speculate that the primary causative factor of the fatigue fracture of the base plate due to loss of bony support most likely secondary to osteolysis was varus malalignment at primary implantation.

This case highlights the importance of paying close attention to the correct alignment of each component at primary implantation.

American Psychological Association (APA)

Mineta, Kazuaki& Okada, Masahiko& Matsumoto, Soshi& Hamada, Daisuke& Goto, Tomohiro& Sairyo, Koichi. 2017. Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty. Case Reports in Orthopedics،Vol. 2017, no. 2017, pp.1-7.
https://search.emarefa.net/detail/BIM-1148291

Modern Language Association (MLA)

Mineta, Kazuaki…[et al.]. Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty. Case Reports in Orthopedics No. 2017 (2017), pp.1-7.
https://search.emarefa.net/detail/BIM-1148291

American Medical Association (AMA)

Mineta, Kazuaki& Okada, Masahiko& Matsumoto, Soshi& Hamada, Daisuke& Goto, Tomohiro& Sairyo, Koichi. Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty. Case Reports in Orthopedics. 2017. Vol. 2017, no. 2017, pp.1-7.
https://search.emarefa.net/detail/BIM-1148291

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1148291