Fracture of a Femoral Revision Stem following a Technical Failure

Joint Authors

Herold, Falko
Eijer, Henk

Source

Case Reports in Orthopedics

Issue

Vol. 2018, Issue 2018 (31 Dec. 2018), pp.1-3, 3 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2018-06-19

Country of Publication

Egypt

No. of Pages

3

Main Subjects

Medicine

Abstract EN

We report about a fracture of a modular, uncemented femoral revision stem (Zimmer, Revitan®) due to a not previously described intraoperative technical problem.

During implantation, a small ring, part of the proximal part of the trial stem, was left on the distal part of the definite stem.

Following this, the top screwcap of the proximal part of the definite stem could not be tightened properly.

However, the stem was thought to be stable, and the situation left.

Two and a half years later, the proximal part of the stem fractured and the situation became unstable.

It is very useful to know about this ring and that it should always be removed, otherwise, there is possibility that it may be left on the definite distal part of the stem with the possibility of a later fracture.

American Psychological Association (APA)

Herold, Falko& Eijer, Henk. 2018. Fracture of a Femoral Revision Stem following a Technical Failure. Case Reports in Orthopedics،Vol. 2018, no. 2018, pp.1-3.
https://search.emarefa.net/detail/BIM-1148295

Modern Language Association (MLA)

Herold, Falko& Eijer, Henk. Fracture of a Femoral Revision Stem following a Technical Failure. Case Reports in Orthopedics No. 2018 (2018), pp.1-3.
https://search.emarefa.net/detail/BIM-1148295

American Medical Association (AMA)

Herold, Falko& Eijer, Henk. Fracture of a Femoral Revision Stem following a Technical Failure. Case Reports in Orthopedics. 2018. Vol. 2018, no. 2018, pp.1-3.
https://search.emarefa.net/detail/BIM-1148295

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1148295