A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode

Joint Authors

Salama, Abdalla
Malekmohammadi, Amin
Mohanna, Shahram
Rajkumar, Rajprasad

Source

International Journal of Biomedical Imaging

Issue

Vol. 2017, Issue 2017 (31 Dec. 2017), pp.1-19, 19 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2017-10-31

Country of Publication

Egypt

No. of Pages

19

Main Subjects

Medicine

Abstract EN

This paper presents a multitasking electrical impedance tomography (EIT) system designed to improve the flexibility and durability of an existing EIT system.

The ability of the present EIT system to detect, locate, and reshape objects was evaluated by four different experiments.

The results of the study show that the system can detect and locate an object with a diameter as small as 1.5 mm in a testing tank with a diameter of 134 mm.

Moreover, the results demonstrate the ability of the current system to reconstruct an image of several dielectric object shapes.

Based on the results of the experiments, the programmable EIT system can adapt the EIT system for different applications without the need to implement a new EIT system, which may help to save time and cost.

The setup for all the experiments consisted of a testing tank with an attached 16-electrode array made of titanium alloy grade 2.

The titanium alloy electrode was used to enhance EIT system’s durability and lifespan.

American Psychological Association (APA)

Salama, Abdalla& Malekmohammadi, Amin& Mohanna, Shahram& Rajkumar, Rajprasad. 2017. A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode. International Journal of Biomedical Imaging،Vol. 2017, no. 2017, pp.1-19.
https://search.emarefa.net/detail/BIM-1159647

Modern Language Association (MLA)

Salama, Abdalla…[et al.]. A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode. International Journal of Biomedical Imaging No. 2017 (2017), pp.1-19.
https://search.emarefa.net/detail/BIM-1159647

American Medical Association (AMA)

Salama, Abdalla& Malekmohammadi, Amin& Mohanna, Shahram& Rajkumar, Rajprasad. A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode. International Journal of Biomedical Imaging. 2017. Vol. 2017, no. 2017, pp.1-19.
https://search.emarefa.net/detail/BIM-1159647

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID