A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode
Joint Authors
Salama, Abdalla
Malekmohammadi, Amin
Mohanna, Shahram
Rajkumar, Rajprasad
Source
International Journal of Biomedical Imaging
Issue
Vol. 2017, Issue 2017 (31 Dec. 2017), pp.1-19, 19 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2017-10-31
Country of Publication
Egypt
No. of Pages
19
Main Subjects
Abstract EN
This paper presents a multitasking electrical impedance tomography (EIT) system designed to improve the flexibility and durability of an existing EIT system.
The ability of the present EIT system to detect, locate, and reshape objects was evaluated by four different experiments.
The results of the study show that the system can detect and locate an object with a diameter as small as 1.5 mm in a testing tank with a diameter of 134 mm.
Moreover, the results demonstrate the ability of the current system to reconstruct an image of several dielectric object shapes.
Based on the results of the experiments, the programmable EIT system can adapt the EIT system for different applications without the need to implement a new EIT system, which may help to save time and cost.
The setup for all the experiments consisted of a testing tank with an attached 16-electrode array made of titanium alloy grade 2.
The titanium alloy electrode was used to enhance EIT system’s durability and lifespan.
American Psychological Association (APA)
Salama, Abdalla& Malekmohammadi, Amin& Mohanna, Shahram& Rajkumar, Rajprasad. 2017. A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode. International Journal of Biomedical Imaging،Vol. 2017, no. 2017, pp.1-19.
https://search.emarefa.net/detail/BIM-1159647
Modern Language Association (MLA)
Salama, Abdalla…[et al.]. A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode. International Journal of Biomedical Imaging No. 2017 (2017), pp.1-19.
https://search.emarefa.net/detail/BIM-1159647
American Medical Association (AMA)
Salama, Abdalla& Malekmohammadi, Amin& Mohanna, Shahram& Rajkumar, Rajprasad. A Multitasking Electrical Impedance Tomography System Using Titanium Alloy Electrode. International Journal of Biomedical Imaging. 2017. Vol. 2017, no. 2017, pp.1-19.
https://search.emarefa.net/detail/BIM-1159647
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID