Design of a Remote Real-Time Monitoring System for Multiple Physiological Parameters Based on Smartphone

Joint Authors

Al-Naggar, Noman Q.
Al-Hammadi, Husam Mohammed
Al-Fusail, Adel Mohammed
AL-Shaebi, Zakarya Ali

Source

Journal of Healthcare Engineering

Issue

Vol. 2019, Issue 2019 (31 Dec. 2019), pp.1-13, 13 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2019-11-19

Country of Publication

Egypt

No. of Pages

13

Main Subjects

Public Health
Medicine

Abstract EN

Background.

Utilization of the widely used wearable sensor and smartphone technology for remote monitoring represents a healthcare breakthrough.

This study aims to design a remote real-time monitoring system for multiple physiological parameters (electrocardiogram, heart rate, respiratory rate, blood oxygen saturation, and temperature) based on smartphones, considering high performance, autoalarm generation, warning transmission, and security through more than one method.

Methods.

Data on monitoring parameters were acquired by the integrated circuits of wearable sensors and collected by an Arduino Mega 250 R3.

The collected data were transmitted via a Wi-Fi interface to a smartphone.

A patient application was developed to analyze, process, and display the data in numerical and graphical forms.

The abnormality threshold values of parameters were identified and analyzed to generate an autoalarm in the system and transmitted with data to a doctor application via a third-generation (3G) mobile network and Wi-Fi.

The performance of the proposed system was verified and evaluated.

The proposed system was designed to meet main (sensing, processing, displaying, real-time transmission, autoalarm generation, and threshold value identification) and auxiliary requirements (compatibility, comfort, low power consumption and cost, small size, and suitability for ambulatory applications).

Results.

System performance is reliable, with a sufficient average accuracy measurement (99.26%).

The system demonstrates an average time delay of 14 s in transmitting data to a doctor application via Wi-Fi compared with an average time of 68 s via a 3G mobile network.

The proposed system achieves low power consumption against time (4 h 21 m 30 s) and the main and auxiliary requirements for remotely monitoring multiple parameters simultaneously with secure data.

Conclusions.

The proposed system can offer economic benefits for remotely monitoring patients living alone or in rural areas, thereby improving medical services, if manufactured in large quantities.

American Psychological Association (APA)

Al-Naggar, Noman Q.& Al-Hammadi, Husam Mohammed& Al-Fusail, Adel Mohammed& AL-Shaebi, Zakarya Ali. 2019. Design of a Remote Real-Time Monitoring System for Multiple Physiological Parameters Based on Smartphone. Journal of Healthcare Engineering،Vol. 2019, no. 2019, pp.1-13.
https://search.emarefa.net/detail/BIM-1175264

Modern Language Association (MLA)

Al-Naggar, Noman Q.…[et al.]. Design of a Remote Real-Time Monitoring System for Multiple Physiological Parameters Based on Smartphone. Journal of Healthcare Engineering No. 2019 (2019), pp.1-13.
https://search.emarefa.net/detail/BIM-1175264

American Medical Association (AMA)

Al-Naggar, Noman Q.& Al-Hammadi, Husam Mohammed& Al-Fusail, Adel Mohammed& AL-Shaebi, Zakarya Ali. Design of a Remote Real-Time Monitoring System for Multiple Physiological Parameters Based on Smartphone. Journal of Healthcare Engineering. 2019. Vol. 2019, no. 2019, pp.1-13.
https://search.emarefa.net/detail/BIM-1175264

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1175264