Novel Passive Two-Stage Magnetic Targeting Devices for Distal Locking of Interlocking Nails

Joint Authors

Wong, Tze-Hong
Lee, Meng-Shiue
Wu, Sung-Yueh
Hsu, Wensyang
Chung, Tien-Kan
Wu, Chia-Pei
Hsu, Pei-Jung
Yang, Yuh-Shyong

Source

Journal of Healthcare Engineering

Issue

Vol. 2017, Issue 2017 (31 Dec. 2017), pp.1-11, 11 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2017-08-13

Country of Publication

Egypt

No. of Pages

11

Main Subjects

Public Health
Medicine

Abstract EN

Interlocking nailing is a common surgical operation to stabilize fractures in long bones.

One of the difficult parts of the surgery is how to locate the position and direction of a screw hole on the interlocking nail, which is invisible to the naked eye after insertion of the nail into the medullary canal.

Here, we propose a novel two-stage targeting process using two passive magnetic devices to locate the position and direction of the screw hole without radiation for the locking screw procedure.

This involves a ring-shape positioning magnet inside the nail to generate a magnetic field for targeting.

From the accuracy test results of these two-stage targeting devices, the search region can be identified in less than 20 seconds by the 1st-stage targeting device, while the total targeting time to locate the drilling position and direction takes less than 4 minutes, with 100% successful rate in 50 attempts.

The drilling test further combines the two-stage targeting process and drilling process on the swine tibia, and it is shown that a 100% successful rate is achieved in all 10 attempts, where the total time needed is less than 5 minutes.

American Psychological Association (APA)

Wong, Tze-Hong& Lee, Meng-Shiue& Wu, Sung-Yueh& Hsu, Wensyang& Chung, Tien-Kan& Wu, Chia-Pei…[et al.]. 2017. Novel Passive Two-Stage Magnetic Targeting Devices for Distal Locking of Interlocking Nails. Journal of Healthcare Engineering،Vol. 2017, no. 2017, pp.1-11.
https://search.emarefa.net/detail/BIM-1180908

Modern Language Association (MLA)

Wong, Tze-Hong…[et al.]. Novel Passive Two-Stage Magnetic Targeting Devices for Distal Locking of Interlocking Nails. Journal of Healthcare Engineering No. 2017 (2017), pp.1-11.
https://search.emarefa.net/detail/BIM-1180908

American Medical Association (AMA)

Wong, Tze-Hong& Lee, Meng-Shiue& Wu, Sung-Yueh& Hsu, Wensyang& Chung, Tien-Kan& Wu, Chia-Pei…[et al.]. Novel Passive Two-Stage Magnetic Targeting Devices for Distal Locking of Interlocking Nails. Journal of Healthcare Engineering. 2017. Vol. 2017, no. 2017, pp.1-11.
https://search.emarefa.net/detail/BIM-1180908

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1180908