Totally Implanted Port May Be an Alternative to Centrally Inserted Central Catheter for Measurement of Central Venous Pressure

Joint Authors

Huang, Chih-Yu
Yu, Chung-Chieh
Kuo, Wei-Ke
Hua, Chung-Ching

Source

Journal of Healthcare Engineering

Issue

Vol. 2020, Issue 2020 (31 Dec. 2020), pp.1-7, 7 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2020-06-30

Country of Publication

Egypt

No. of Pages

7

Main Subjects

Public Health
Medicine

Abstract EN

Background.

A conventional centrally inserted central catheter (CICC) is frequently used to measure central venous pressure (CVP) to monitor the cardiocirculatory status of patients.

The tip of the totally implanted port (TIP) is inserted at the same location in the superior vena cava as that of a CICC, and the TIP has been implanted in many patients with cancer.

Measurements of CVP using CICC (CICCP) and TIP (TIPP) may be closely related.

Material and Methods.

Ten patients with TIPs in an intensive care unit were prospectively studied, and 121 records of 4536 paired CICCP and TIPP measurements were collected.

A bench test in a static or dynamic setting was performed, and 598 paired measurements taken using CICC and TIP were recorded.

Results.

The measurement of TIPP was highly correlated with that of CICCP in patients with cancer, especially those in a calm state.

Patients with a calm state and ≥3 consecutive identical TIPP were recorded (≥30 seconds), and 90% of the mean difference between CICCP and TIPP was ≤2 mmHg.

The pressure measurements recorded using CICC and TIP were identical in both the static and dynamic bench tests.

Conclusions.

TIP may be an alternative to CICC for measuring CVP.

American Psychological Association (APA)

Kuo, Wei-Ke& Huang, Chih-Yu& Yu, Chung-Chieh& Hua, Chung-Ching. 2020. Totally Implanted Port May Be an Alternative to Centrally Inserted Central Catheter for Measurement of Central Venous Pressure. Journal of Healthcare Engineering،Vol. 2020, no. 2020, pp.1-7.
https://search.emarefa.net/detail/BIM-1186710

Modern Language Association (MLA)

Kuo, Wei-Ke…[et al.]. Totally Implanted Port May Be an Alternative to Centrally Inserted Central Catheter for Measurement of Central Venous Pressure. Journal of Healthcare Engineering No. 2020 (2020), pp.1-7.
https://search.emarefa.net/detail/BIM-1186710

American Medical Association (AMA)

Kuo, Wei-Ke& Huang, Chih-Yu& Yu, Chung-Chieh& Hua, Chung-Ching. Totally Implanted Port May Be an Alternative to Centrally Inserted Central Catheter for Measurement of Central Venous Pressure. Journal of Healthcare Engineering. 2020. Vol. 2020, no. 2020, pp.1-7.
https://search.emarefa.net/detail/BIM-1186710

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-1186710