Vital-SCOPE: Design and Evaluation of a Smart Vital Sign Monitor for Simultaneous Measurement of Pulse Rate, Respiratory Rate, and Body Temperature for Patient Monitoring

Joint Authors

Sun, Guanghao
Matsui, Takemi
Watai, Yasuyuki
Kim, Seokjin
Kirimoto, Tetsuo
Suzuki, Satoshi
Hakozaki, Yukiya

Source

Journal of Sensors

Issue

Vol. 2018, Issue 2018 (31 Dec. 2018), pp.1-7, 7 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2018-02-01

Country of Publication

Egypt

No. of Pages

7

Main Subjects

Civil Engineering

Abstract EN

Consistent vital sign monitoring is critically important for early detection of clinical deterioration of patients in hospital settings.

Mostly, nurses routinely measure and document the primary vital signs of all patients 2‐3 times daily to assess their condition.

To reduce nurse workload and thereby improve quality of patient care, a smart vital sign monitor named “Vital‐SCOPE” for simultaneous measurement of vital signs was developed.

Vital-SCOPE consists of multiple sensors, including a reflective photo sensor, thermopile, and medical radar, to be used in simultaneous pulse rate, respiratory rate, and body temperature monitoring within 10 s.

It was tested in laboratory and hospital settings.

Bland-Altman and Pearson’s correlation analyses were used to compare the Vital-SCOPE results to those of reference measurements.

The mean difference of the respiratory rate between respiratory effort belt and Vital-SCOPE was 0.47 breaths per minute with the 95% limit of agreement ranging from −7.4 to 6.5 breaths per minute.

The Pearson’s correlation coefficient was 0.63 (P<0.05).

Moreover, the mean difference of the pulse rate between electrocardiogram and Vital-SCOPE was 3.4 beats per minute with the 95% limit of agreement ranging from −13 to 5.8 beats per minute; the Pearson’s correlation coefficient was 0.91 (P<0.01), indicating strong linear relationship.

American Psychological Association (APA)

Sun, Guanghao& Matsui, Takemi& Watai, Yasuyuki& Kim, Seokjin& Kirimoto, Tetsuo& Suzuki, Satoshi…[et al.]. 2018. Vital-SCOPE: Design and Evaluation of a Smart Vital Sign Monitor for Simultaneous Measurement of Pulse Rate, Respiratory Rate, and Body Temperature for Patient Monitoring. Journal of Sensors،Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1201468

Modern Language Association (MLA)

Sun, Guanghao…[et al.]. Vital-SCOPE: Design and Evaluation of a Smart Vital Sign Monitor for Simultaneous Measurement of Pulse Rate, Respiratory Rate, and Body Temperature for Patient Monitoring. Journal of Sensors No. 2018 (2018), pp.1-7.
https://search.emarefa.net/detail/BIM-1201468

American Medical Association (AMA)

Sun, Guanghao& Matsui, Takemi& Watai, Yasuyuki& Kim, Seokjin& Kirimoto, Tetsuo& Suzuki, Satoshi…[et al.]. Vital-SCOPE: Design and Evaluation of a Smart Vital Sign Monitor for Simultaneous Measurement of Pulse Rate, Respiratory Rate, and Body Temperature for Patient Monitoring. Journal of Sensors. 2018. Vol. 2018, no. 2018, pp.1-7.
https://search.emarefa.net/detail/BIM-1201468

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID