Analysis of a Novel Magnetic-Hydrodynamic Double Levitated Motor for an Implantable Axial Flow Blood Pump
Joint Authors
Xiang, Chuang
Wang, Liang
Tang, Xiaoyan
Yun, Zhong
Source
Issue
Vol. 2020, Issue 2020 (31 Dec. 2020), pp.1-15, 15 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2020-12-10
Country of Publication
Egypt
No. of Pages
15
Main Subjects
Abstract EN
This paper presents a novel design for a bearingless axial flow blood pump based on the magnetic-hydrodynamic double levitated concept.
In the axial direction, the magnetic levitation system consisted of two pairs of permanent magnet rings offsets the force of fluid.
The hydrodynamic shell mounted on the impeller rotor is designed for generating dynamic pressure, which can balance the radial force like gravity when the blood pump is working.
Because of the unsteady force and torque acting on the rotor and the passive suspension, the position of the rotor is not steady.
The suspension force, stiffness, and torque of the rotor are calculated by the theoretical method and finite element method.
Then, the dynamics of the rotor are analyzed.
Arrangements of Hall-effect sensors with the corresponding data acquisition system which can measure the axial displacement of the rotor are explained.
The sensorless drive control system for the blood pump is described too.
With a prototype pump, an external circulation experiment system is built and then the axial and radial displacements of the rotor are measured by using Hall-effect sensors and the laser vibrometer under different working conditions.
American Psychological Association (APA)
Wang, Liang& Tang, Xiaoyan& Yun, Zhong& Xiang, Chuang. 2020. Analysis of a Novel Magnetic-Hydrodynamic Double Levitated Motor for an Implantable Axial Flow Blood Pump. Shock and Vibration،Vol. 2020, no. 2020, pp.1-15.
https://search.emarefa.net/detail/BIM-1212808
Modern Language Association (MLA)
Wang, Liang…[et al.]. Analysis of a Novel Magnetic-Hydrodynamic Double Levitated Motor for an Implantable Axial Flow Blood Pump. Shock and Vibration No. 2020 (2020), pp.1-15.
https://search.emarefa.net/detail/BIM-1212808
American Medical Association (AMA)
Wang, Liang& Tang, Xiaoyan& Yun, Zhong& Xiang, Chuang. Analysis of a Novel Magnetic-Hydrodynamic Double Levitated Motor for an Implantable Axial Flow Blood Pump. Shock and Vibration. 2020. Vol. 2020, no. 2020, pp.1-15.
https://search.emarefa.net/detail/BIM-1212808
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-1212808