Comparison of hearing threshold levels using tdh-39 and tdh-49 earphones
Joint Authors
al-Tamimi, Shawkat F.
al-Qasim, Husayn Y.
al-Jundi, Ali M.
Source
Journal of the Royal Medical Services
Issue
Vol. 9, Issue 1 (30 Jun. 2002), pp.75-78, 4 p.
Publisher
The Royal Medical Services Jordan Armed Forces
Publication Date
2002-06-30
Country of Publication
Jordan
No. of Pages
4
Main Subjects
Topics
Abstract EN
Objective : The study compares hearing threshold levels obtained on the same ear, by using four types of earphones with particular emphasis on the region of 6 kHz, to discover how the differences measured on the coupler translate to measure hearing threshold levels.
Methods : Air conduction audiometry on the same ear of 20 subjects of both sexes using the manual technique was undertaken using four types of earphones, all fitted with MX- 41 / AR cushion using a clinical audiometer type KcSO using 2 dB steps.
Results : The human results are consistent with previous coupler results, showing differences in the order of 6 dB around 6 kHz using international standard methods for calibration; that is using the IEC 303 coupler and associated reference equivalent threshold sound pressure levels for the TDH-39 and the IEC 318 artificial ear for the TDH-49 and TDH-49P.
The source of discrepancy appears to be a specific interaction between the TDH-39 earphone and the IEC 303 coupler. Conclusion : It is concluded that a move away from the use of two different couplers and two different sets of associated reference equivalent threshold sound pressure levels for different earphone types would be beneficial to avoid the problem of differing performance.
Moreover, the IEC 318 artificial ear is recognized as being a better approximation at the human ear than the IEC 303 coupler.
American Psychological Association (APA)
al-Tamimi, Shawkat F.& al-Jundi, Ali M.& al-Qasim, Husayn Y.. 2002. Comparison of hearing threshold levels using tdh-39 and tdh-49 earphones. Journal of the Royal Medical Services،Vol. 9, no. 1, pp.75-78.
https://search.emarefa.net/detail/BIM-129960
Modern Language Association (MLA)
al-Tamimi, Shawkat F.…[et al.]. Comparison of hearing threshold levels using tdh-39 and tdh-49 earphones. Journal of the Royal Medical Services Vol. 9, no. 1 (Jun. 2002), pp.75-78.
https://search.emarefa.net/detail/BIM-129960
American Medical Association (AMA)
al-Tamimi, Shawkat F.& al-Jundi, Ali M.& al-Qasim, Husayn Y.. Comparison of hearing threshold levels using tdh-39 and tdh-49 earphones. Journal of the Royal Medical Services. 2002. Vol. 9, no. 1, pp.75-78.
https://search.emarefa.net/detail/BIM-129960
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references : p. 77-78
Record ID