![](/images/graphics-bg.png)
Investigation of helical strut attached vena cava filter hemodynamic performance
Author
Source
Journal of Engineering Research
Issue
Vol. 10, Issue 2 B (30 Jun. 2022), pp.174-183, 10 p.
Publisher
Kuwait University Academic Publication Council
Publication Date
2022-06-30
Country of Publication
Kuwait
No. of Pages
10
Main Subjects
Abstract EN
Hemodynamic performance of the celect platinum vena cava filter and the revised forms of it with helical flow inducer strut were studied with computational fluid dynamic software ansys fluent 18.
the central velocity and shear stress increased but overall flow disturbance has been observed minimal level.
central velocity increases to 9.72% with celect filter, by the single helical flow inducer strut the rate reaches to 14.69%, and with doubled form it reaches to 19.73%.
the filter surface shear rate increases to 8.29% with the single helical flow inducer strut and increases 13.31% with doubled attachment.
increased velocity and shear stress on the filter may eliminate short term thrombus build-up problems by breaking the big size particulates with the high shearing forces.
the new struts may also contribute to the ability of the filter to capture smaller clots, as well as to dissolve them from being bigger.
shear stress in the vein wall increases approximately 6.63% with the filter placement.
it raises to 8.06% and 9.45% with single and double helical flow inducer strut attachment.
increased vein wall shear may reduce the recirculation and clotting in the vein wall and it may prevent the accumulation of clots.
the increased shear stress on the filter may cause the migration problem, design improvements can minimize this risk.
helical flow inducer strut attachment can cause efficacy increase, and the flow are normalized.
American Psychological Association (APA)
Selimli, Selcuk. 2022. Investigation of helical strut attached vena cava filter hemodynamic performance. Journal of Engineering Research،Vol. 10, no. 2 B, pp.174-183.
https://search.emarefa.net/detail/BIM-1494658
Modern Language Association (MLA)
Selimli, Selcuk. Investigation of helical strut attached vena cava filter hemodynamic performance. Journal of Engineering Research Vol. 10, no. 2 B (Jun. 2022), pp.174-183.
https://search.emarefa.net/detail/BIM-1494658
American Medical Association (AMA)
Selimli, Selcuk. Investigation of helical strut attached vena cava filter hemodynamic performance. Journal of Engineering Research. 2022. Vol. 10, no. 2 B, pp.174-183.
https://search.emarefa.net/detail/BIM-1494658
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references : p. 182-183
Record ID