![](/images/graphics-bg.png)
Development of high-performance liquid chromatographic method for vitamin d3 analysis in pharmaceutical preparation
Other Title(s)
تطوير طريقة لتحليل فيتامين ( د 3 ) في جهاز الكروماتوغرافيا السائلة ذات الكفاءة العالية في المستحضرات الصيدلانية
Joint Authors
al-Hadidi, Kamal
al-Qadi, Inas
Battah, Abd al-Qadir
Source
Jordan Journal of Pharmaceutical Sciences
Issue
Vol. 3, Issue 2 (31 Jul. 2010), pp.78-86, 9 p.
Publisher
University of Jordan Deanship of Academic Research (DAR)
Publication Date
2010-07-31
Country of Publication
Jordan
No. of Pages
9
Main Subjects
Pharmacy, Health & Medical Sciences
Abstract AR
توصلت الدراسة إلى تطوير طريقة سهلة، و سريعة، و اقتصادية لتحليل فيتامين د في المستحضرات الصيدلانية بواسطة الطور الثابت (C18, 150 X 4.6 mm, 3.5µm).
كما تم استخدام الميثانول و الاسيتونيترال بنسبة 75 : 25 كطور متحرك بنسبة تدفق تساوي 1 مليلتر لكل دقيقة عند درجة حرارة 40 مْ، و درجة امتصاص 265 نانو ميتر خلال فترة زمنية تقل عن 12 دقيقة.
و قد تم الحصول على نتائج مرضية لعملية تقييم دقة الطريقة المطورة و ضبطها، إذ تراوحت نتائج حساب دقة الطريقة المطورة ما بين 6.267 ± 0.050 – 2.600 ± 0.078 % %.
كما تم اختبار ضبط الطريقة المطورة من خلال حساب معامل التغير خلال يوم و عدة أيام و كانت النتيجة ما بين 0.050 ± 0.001% - 2.144% ± 0.002.
كما كان خط المعايرة مستقيمًا لكل من التراكيز 0 إلى 10 مايكرو جرام / مليلتر لفيتامين د 3 .
و نسبة الشوائب تساوي 3أو أقل.
كما أن الفيتامينات: أ، وه، وه أسيتيت لم تتداخل أو تتأثر مع فيتامين د.
و قد تم تطبيق هذه الطريقة المطورة على ست عشرة عينة صيدلانية صلبة تحتوي على فيتامين د مع الكالسيوم، قد أظهرت النتائج أن 31.25 % من العينات المختبرة أعطت قيمًا موافقة لتلك القيم المصرح عنها في تلك الأدوية .
Abstract EN
In the present study, a rapid, simple and economical reversed phase-HPLC procedure has been developed for the determination of vitamin D3 and vitamin D2 in pharmaceutical preparations.
The vitamins were separated isocratically on a Teknokroma C18 column (150×4.6mm, 3.5μm particle size) with a mobile phase consisting of isocratic acetonitrile and methanol (75:25, v/v) operated at 40ºC with retention times less than 12 min.
The eluted vitamins were identified and monitored on a diode-array detector (DAD) at 265nm.
The linearity of the method was excellent (r2 > 0.999), over the concentration range of 0-10μg/ml.
The statistical evaluation of the method was examined performing intra-day and inter-day calibrations and were found to be satisfactory, with 2.600 ± 0.078% and 6.267 ± 0.050% accuracy and 0.001 ± 0.002% to 2.144 ± 0.050% precision results.
Mean recovery of vitamin D3 was 98.2 ± 0.349% for different spiking levels.
Detection limit was found to be 25ng/ml.
No interference was found from other fat soluble vitamins (vitamin A, vitamin E and vitamin E acetate) that are commonly presents with vitamin D pharmaceutical preparation.
The method was applied for the analysis of sixteen solid samples of pharmaceutical preparation (calcium with vitamin D).
The results show only 31.25% of the tested samples give satisfactory agreement with the declared values.
Keywords: Vit.
D, HPLC, DAD and Pharmaceutical Preparation.
American Psychological Association (APA)
al-Qadi, Inas& Battah, Abd al-Qadir& al-Hadidi, Kamal. 2010. Development of high-performance liquid chromatographic method for vitamin d3 analysis in pharmaceutical preparation. Jordan Journal of Pharmaceutical Sciences،Vol. 3, no. 2, pp.78-86.
https://search.emarefa.net/detail/BIM-272683
Modern Language Association (MLA)
al-Qadi, Inas…[et al.]. Development of high-performance liquid chromatographic method for vitamin d3 analysis in pharmaceutical preparation. Jordan Journal of Pharmaceutical Sciences Vol. 3, no. 2 (Jul. 2010), pp.78-86.
https://search.emarefa.net/detail/BIM-272683
American Medical Association (AMA)
al-Qadi, Inas& Battah, Abd al-Qadir& al-Hadidi, Kamal. Development of high-performance liquid chromatographic method for vitamin d3 analysis in pharmaceutical preparation. Jordan Journal of Pharmaceutical Sciences. 2010. Vol. 3, no. 2, pp.78-86.
https://search.emarefa.net/detail/BIM-272683
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references : p. 85
Record ID
BIM-272683