The determination of 25-OHVitamin (D2 D3)‎ in human serum by liquid chromatography tandem mass spectrometry with comparison to IDS enzyme immunoassay

Other Title(s)

تحديد فيتامين د 2 و د 3 (هيدروكسي 25)‎ في مصل الدم البشري باستخدام تقنية الكراماتغرافيا السائلة مقرونة بقياس الطيف الكتلي (ID-LEMSTMS)‎

Joint Authors

Milhim, Samar J.
Aidah, Khalid M. A.
al-Hadidi, Kamal A.
al-Alali, Faris

Source

Jordan Journal of Pharmaceutical Sciences

Issue

Vol. 6, Issue 2 (31 May. 2013), pp.203-222, 20 p.

Publisher

University of Jordan Deanship of Academic Research (DAR)

Publication Date

2013-05-31

Country of Publication

Jordan

No. of Pages

20

Main Subjects

Pharmacy, Health & Medical Sciences

Topics

Abstract AR

تم استخدام تقنية الكرماتغرافيا السائلة مقرونة بقياس الطيف الكتلي (ID-LC-MS / MS) لقياس نسبة 25 هيدروكسي فيتامين د2 و 25 هيدروكسي فيتامين د3 في مصل الدم البشري.

أضيف سداسي الديتروم 25 هيدروكسي فيتامين د3 كمادة معيارية داخلية لأنظمة التدريج (كاليبريتور) و التحكم (كونترول)-المحضرة في زلال مصل بقري نسبته 6 % و ملح الفوسفات كذلك أضيف إلى عينات أمصال المرضى.

تمت عملية الفصل الكرمتغرافي بتقنية أولية عن طريقة عمود فخ (trap column) يليه الفصل المروماتوغرافي على عمود C-8 باستخدام تدرج مائي / الأسيتونيتريل.

وقد أجري الكشف عن النواتج الكيميائية بواسطة نظام التأين الكيميائي ذي الضغط الجوي (AP-CI) في رصد ردود الفعل متعددة (MRM) الواسطة.

و كانت طريقة الرصد ذات نمط خطي واحد يتراوح ما بين 4 إلى 70 نانوغرام / مليلتر.

بلغت النسبة المئوية لمعامل التباين (CV %) خلال اليوم الواحد و خلال عدة أيام ≤ 10 % لكل من 25-هيدروكسي فيتامين د2 و 25-هيدروكسي فيتامين د3.

تراوح الناتج المسترد ما بين 39.09 % إلى 64.31 % ل 25- فيتامين د2 و 30.44 % فيتامين د3 إلى 58.66 % ل 25- هيدروكسي فيتامين د3 بينما تراوح الناتج المسترد لسداسي الديتيريوم 25-هيدروكسي فيتامين د3 بين 44.11 % إلى 67.5 %.

تمت مقارنة طريقة الكرماتغرافيا السائلة المقرونة بمقياس الطيف الكتلي الميتحدثة بتقنية الأنزيم المناعي من شركة (IDS) Immunodiagnostic system من حيث التوافق و الارتباط و تأثير اختلاف طريقة الفحص على التصنيف التشخيصي لحالة فيتامين د بواسطة قياس نسبته في عينات 182 من دم المرضى بطريقة مقارنة ؛ كانت نسبة التحيز (Bias %) لتقنية الإنزيم المناعي مقارنة بطريقة الكرماتغرافيا السائلة 10.15 ± 34.28-بفترة ثقة احتمالية 95 % 24.13] إلى [44.43 وجد أن هناك توافقا جيدا مع ارتباط معقول (0.82 = r2) ,(P value = 0.000) كان التصنيف التشخيصي بين الطريقتين متغيرا، و يرتكز بشكل أساسي على نوع الطريقة المستخدمة و القيم الفاصلة المتبعة لتعريف النقص في فيتامين د، لذلك فإن توحيد طرق فحص فيتامين د بطرق معيارية شاملة ضروري جدا من أجل التقييم السريري السليم.

تعتبر طريقة الكرمتغرافيا السائلة المستحدثة متينة من الناحية التحليلية و موثوقا بها مخبريا لقياس كل من 25-هيدروكسي فيتامين د2 و 25-هيدروكسي فيتامين د3 في مصل الدم البشري.

Abstract EN

The proper assessment of the status of vitamin D requires the accurate measurement of both 25-OH vitamin D2 and 25-OH vitamin D3, which collectively constitute 25-OH vitamin D, the best indicator of vitamin D status.

Currently, numerous assay methods are available for 25-OH vitamin D measurement but their comparability is uncertain.

We employed isotope dilution liquid chromatography coupled with tandem mass spectrometry (ID-LC-MS / MS) to quantify 25-OH vitamin D2 and 25-OH vitamin D3 in human serum.

Hex deuterium labeled 25-OH vitamin D3 internal standard was added to calibrators, controls prepared in 6 % bovine serum albumin in phosphate buffered saline, and patients’ sera.

Zinc sulphate was added to release 25-OH vitamin D metabolites for vitamin D binding protein, followed by a precipitation step with the addition of acetonitrile.

Subsequent online phase extraction by trap column followed by chromatographic separation on a C-8 column using a water / acetonitrile gradient was employed.

Detection was performed using Atmospheric Pressure Chemical Ionization (AP-CI) in a Multiple Reaction Monitoring (MRM) mode.

The method was linear from 4 to 70 ng / mL.

The intra and inter-day CV % were ≤ 10 for both 25-OH vitamin D2 and 25-OH vitamin D3.

Recoveries ranged between 39.09 % to 64.31 % for 25-OH vitamin D2 and 30.44 % to 58.66 % for 25-OH vitamin D3, while recoveries from hexadeuterium 25-OH vitamin D3 ranged from 44.11 % to 67.5 %.We compared the newly validated LC-MS / MS with a commercial Enzyme Immunoassay from Immunodiagnostic Systems (IDS EIA) in terms of inter-method agreement, correlation, and impact of assay variation on the diagnostic categorization of vitamin D status through the measurement of 182 subjects’ sera.

The mean bias % of the IDS EIA relative to the LC-MS / MS was -34.28 ± (10.15 mean ± std) with 95 % CI [-24.13 to 44.43].

The two methods were in good agreement with reasonable correlation (r² = 0.82, P value = 0.000).

Inter-method diagnostic categorization was variable and depended on the type of assay method and the applied cut offs used.

Cross calibration and standardization of vitamin D assay methods is crucial for proper clinical assessment of vitamin D status.

This LC-MS / MS method is reliable and robust for the measurement of both 25-OH vitamin D2 and 25-OH vitamin D3 in human serum.

American Psychological Association (APA)

Milhim, Samar J.& Aidah, Khalid M. A.& al-Hadidi, Kamal A.& al-Alali, Faris. 2013. The determination of 25-OHVitamin (D2 D3) in human serum by liquid chromatography tandem mass spectrometry with comparison to IDS enzyme immunoassay. Jordan Journal of Pharmaceutical Sciences،Vol. 6, no. 2, pp.203-222.
https://search.emarefa.net/detail/BIM-326181

Modern Language Association (MLA)

Milhim, Samar J.…[et al.]. The determination of 25-OHVitamin (D2 D3) in human serum by liquid chromatography tandem mass spectrometry with comparison to IDS enzyme immunoassay. Jordan Journal of Pharmaceutical Sciences Vol. 6, no. 2 (2013), pp.203-222.
https://search.emarefa.net/detail/BIM-326181

American Medical Association (AMA)

Milhim, Samar J.& Aidah, Khalid M. A.& al-Hadidi, Kamal A.& al-Alali, Faris. The determination of 25-OHVitamin (D2 D3) in human serum by liquid chromatography tandem mass spectrometry with comparison to IDS enzyme immunoassay. Jordan Journal of Pharmaceutical Sciences. 2013. Vol. 6, no. 2, pp.203-222.
https://search.emarefa.net/detail/BIM-326181

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references : p. 218-221

Record ID

BIM-326181