Comparability study between digital and visual colorimeter by using of chlorine indicators of DpD No.1 and DpD No.3 under practical conditions

Other Title(s)

دراسة مقارنة بين جهاز قياس الكلور الرقمي و جهاز قياس الكلور العيني بواسطة استخدام كواشف الكلور الدي بي دي رقم 1 و الدي بي دي رقم 3 تحت ظروف قياسية

Author

Muhsin, Fulath Abd al-Rida

Source

al-Kufa University Journal for Biology

Issue

Vol. 6, Issue 1 (30 Jun. 2014), pp.1-13, 13 p.

Publisher

University of Kufa Faculty of Science Department of live Sciences

Publication Date

2014-06-30

Country of Publication

Iraq

No. of Pages

13

Main Subjects

Public Health

Topics

Abstract AR

إن كل منظومة ماء شرب بحاجة لقياس مستويات الكلورين المتبقي بأجهزة دقيقة عمليا للتأكد من أن الماء تم تنقيته بصورة مضبوطة.

أجريت هذه الدراسة التجريبية للمقارنة تحت ظروف قياسية في المختبر نتائج فحوصات الكلور الحر و الكلي في نماذج الماء بين جهاز قياس الكلور الرقمي و جهاز قياس الكلور العيني بواسطة استخدام الكواشف الكيمياوية ((الدي بي دي رقم (1)) لقياس الكلورين الحر في نماذج الماء، و استخدام ((الدي بي دي رقم (3)) لقياس الكلورين الكلي في نماذج المياه.

استمرت هذه الدراسة من (1 / 10 / 2011 إلى 1 / 5 / 2012).

تمت الدراسة في مختبرات المعهد التقني الكوفة و التابع لمحافظة النجف الأشرف حوالي 165 كم للجنوب من بغداد، و 60 كم للجنوب من محافظة بابل.

200 تجربة قد تم إجرائها في المختبر، بواسطة العمل بتجربتين في كل يوم تحت ظروف قياسية في المختبر.

كان هنالك فروق معنوية إحصائية كبيرة (0.000 = p) عند المقارنة بين جهاز قياس الكلور الرقمي مع جهاز قياس الكلور العيني لكل من فحوصات الكلور الحر و الكلور الكلي.

كان هنالك فرق كبير في المتوسط الحسابي عند المقارنة بين الجهازين بالنسبة لكل فحوصات الكور الحر و الكلي.

كان هنالك ارتباط موجب و قوي بين نتائج تجارب جهاز الكلور الرقمي و وقت التلامس لكل من فحوصات الكلور الحر و الكلي، بينما كان هنالك ارتباط موجب و ضعيف بين نتائج تجارب جهاز قياس الكلور العيني و وقت التلامس لكل من فحوصات الكلور الحر و الكلي.

من جميع نتائج التجارب المعمولة و المستخلصة من هذه الدراسة، توصي الدراسة من تبني الاستخدام العملي لجهاز قياس الكلور الرقمي بدلا من جهاز قياس الكلور العيني في جميع فحوصات الكلور المتبقي من قبل المفتشين التابعين لوزارة الصحة في جميع المحافظات عند مراقبة مياه الشرب.

كما توصي الدراسة لنتائج أكثر دقة و حساسية من تبني استخدام جهاز قياس الكلور الرقمي في جميع البحوث للتقصي عن الكلور الحر و المتبقي في مختلف نماذج المياه.

Abstract EN

All public drinking water system was required to measure the levels of residual chlorine in a practically accurately instruments to ensure that the water was adequately disinfected.

This experimental study was carried out to compare under practical condition in laboratory the test results of free and total chlorine in the water samples between digital colorimeter and visual colorimeter by using of chemical indicator reagents of ((DpD number (1)) for testing of free chlorine in water samples, and ((DpD number (3)) for testing of total chlorine in water samples.

The experiments lasted from (1 / 10 / 2011 to 1 / 5 / 2012).

The study was conducted in the Laboratories of Al-Kufa Technical Institute in Al-Kufa city that belong to Al-Najaf governorate about 165 Km to the south of Baghdad, and 60 km to the south of Babylon governorate.

200 experiments were done in the laboratory, by made of 2 experiments in each day through a standards laboratory conditions.

There was a highly statistical significant difference (p = 0.000), in comparing between digital calorimeter and visual colorimeter of both free and a total chlorine test results.

There was a highly differences in the mean when was comparison of the digital colorimeter with the visual colorimeter of both all free and total chlorine experiment.

There was a strong direct correlation between digital colorimeter experimental results and contact time in both all free and total chlorine experiments, while there was a weak direct correlation between visual colorimeter and contact time in all free and total chlorine experiments.

From the all tested results of experiment that was derived from this study, the study recommend to practically adaptation of a digital colorimeter instead of visual colorimeter in all the chlorine residual test results that was done by ministry of health inspectors in all governorates in monitoring of drinking water, and for more accurate and sensitive test result we recommend to adapted of digital colorimeter in all researches when tracing of chlorine residual of different water samples.

American Psychological Association (APA)

Muhsin, Fulath Abd al-Rida. 2014. Comparability study between digital and visual colorimeter by using of chlorine indicators of DpD No.1 and DpD No.3 under practical conditions. al-Kufa University Journal for Biology،Vol. 6, no. 1, pp.1-13.
https://search.emarefa.net/detail/BIM-369790

Modern Language Association (MLA)

Muhsin, Fulath Abd al-Rida. Comparability study between digital and visual colorimeter by using of chlorine indicators of DpD No.1 and DpD No.3 under practical conditions. al-Kufa University Journal for Biology Vol. 6, no. 1 (2014), pp.1-13.
https://search.emarefa.net/detail/BIM-369790

American Medical Association (AMA)

Muhsin, Fulath Abd al-Rida. Comparability study between digital and visual colorimeter by using of chlorine indicators of DpD No.1 and DpD No.3 under practical conditions. al-Kufa University Journal for Biology. 2014. Vol. 6, no. 1, pp.1-13.
https://search.emarefa.net/detail/BIM-369790

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references : p. 12-13

Record ID