A new mode for an on-line determination of amiloride in pure and pharmaceutical preparation using CFIA with homemade ayah 3sbgrx3-3d solar cell microphtometer analyzer

Other Title(s)

نمط جديد للتقدير الأني للأميلورايد في المستحضرات الصيدلانية و النقية باستخدام التحميل بالحقن الجرياني المستمر عن طريق استخدام محمل طيفي مايكروي مصنع محميا Ayah 3SBGR x3-3D solar cell.

Joint Authors

al-Awwadi, Nagham Shakir Turki
al-Amir, Malik Husayn al-Allush

Source

Iraqi Journal of Science

Issue

Vol. 55, Issue 3B (31 Dec. 2014), pp.1137-1152, 16 p.

Publisher

University of Baghdad College of Science

Publication Date

2014-12-31

Country of Publication

Iraq

No. of Pages

16

Main Subjects

Chemistry

Topics

Abstract AR

طورت طريقة طيفية جديدة، بسيطة، سريعة، مضبوطة، و حساسة لتحليل الاميلورايد في صيغته النقية و المستحضرات الصيدلانية باستخدام التحليل بالحقن الجرياني المستمر.

استندت الطريقة على أكسدة العقار بكبريتات السيريوم (VI) في الوسط ألحامضي لتكوين أصناف ملونة وردي-محمر و التي قدرت باستخدام مطياف الحقن الجرياني مع Ayah 3S BGR x3-3D solar cell المصنع محليا.

تم دراسة الظروف الفضلى التي تم التوصل إليها باستخدام ثنائي الوصلة باعث للضوء الأخضر و بشدة عالية كمصدر للتشعيع.

كان المدى الخطي الفعال لعلاقة تغير الاستجابة الآلية مع تركيز الاميلورايد 0.0005-10 مللي مول.

لتر-1 بينما حدود الكشف 9.471 نانوغم \ أنموذج.

و أن معامل الارتباط (r) 0.9824 بينما نسبة الخطية r2 % = 96.50 %.

الانحراف القياسي النسبي المؤي (RSD %) للتكرارية أقل من 1.5 % لمحلول الاميلورايد بتركيز 0.5 مللي مول.

لتر-1 (8=n) طبقت الطريقة بنجاح لتقدير الاميلورايد في المستحضرات الصيدلانية.

أجريت مقارنة بين الطريقة المطورة و الطريقة التقليدية (مطيافية UV-Vis) عند طول موجي 540 نانو متر للتحليل باستخدام منحنى الإضافات القياس من خلال استخدام اختبار –t المزدوج.

تبين أنه لا يوجد فرق جوهري بين القيمة الرسمية و قيمة –t المحسوبة من الطريقة المطورة عند فاصل ثقة 95 % بالإضافة إلى إجراء مقارنة بين الطريقتين و حساب قيمة –t و لوحظ فرق جوهري بين الطريقتين عند مستوى قناعة 99 % و فرق جوهري عند قناعة 95 %.

Abstract EN

-A new, simple, accurate, fast and sensitive spectrophotometric method has been developed for the analysis of amiloride in pure and pharmaceutical formulations by continuous flow injection analysis.

The method was based on the oxidation of the drug with Ce(IV)sulfate in acidic medium to formed a pinkish-red color species which determined using homemade Ayah 3SBGR x3-3D solar cell flow injection photometer.

Optimum conditions were obtained using a high intensity green light emitted diode as a source.

The linear dynamic range for the instrument response versus amiloride concentration was 0.0005- 10 mmol.L-1 while the L.O.D was of 9.471 ng / sample.

The correlation coefficient (r) was 0.9824 while percentage linearity (r2%) was 96.50%.

RSD% for the repeatability (n=8) was lower than 1.5% for the determination of amiloride, at concentration of 0.5 mmol.L-1.

The method was applied successfully for the determination of amiloride in pharmaceutical preparation.

A comparison was made between the newly developed method and the classical method (UV-Vis spectrophotometry at wave length 540nm) of analysis using the standard additions method via the use of paired t-test.

It showed that there was no significant difference between the quoted value of each individual company with calculated t-value at 95% confidence interval from developed method, in addition to comparison between two methods and calculate t-value, it was noticed that there no significant difference between two methods at 99% but a significant difference at 95% confidence level.

Key words: Amiloride, flow injection

American Psychological Association (APA)

al-Awwadi, Nagham Shakir Turki& al-Amir, Malik Husayn al-Allush. 2014. A new mode for an on-line determination of amiloride in pure and pharmaceutical preparation using CFIA with homemade ayah 3sbgrx3-3d solar cell microphtometer analyzer. Iraqi Journal of Science،Vol. 55, no. 3B, pp.1137-1152.
https://search.emarefa.net/detail/BIM-394683

Modern Language Association (MLA)

al-Awwadi, Nagham Shakir Turki& al-Amir, Malik Husayn al-Allush. A new mode for an on-line determination of amiloride in pure and pharmaceutical preparation using CFIA with homemade ayah 3sbgrx3-3d solar cell microphtometer analyzer. Iraqi Journal of Science Vol. 55, no. 3B (2014), pp.1137-1152.
https://search.emarefa.net/detail/BIM-394683

American Medical Association (AMA)

al-Awwadi, Nagham Shakir Turki& al-Amir, Malik Husayn al-Allush. A new mode for an on-line determination of amiloride in pure and pharmaceutical preparation using CFIA with homemade ayah 3sbgrx3-3d solar cell microphtometer analyzer. Iraqi Journal of Science. 2014. Vol. 55, no. 3B, pp.1137-1152.
https://search.emarefa.net/detail/BIM-394683

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references : p. 1151-1152

Record ID