Bioanalytical Method Development and Validation of Memantine in Human Plasma by High Performance Liquid Chromatography with Tandem Mass Spectrometry : Application to Bioequivalence Study

Joint Authors

Chandrasekhar, Kothapalli B.
Chandu, Babu Rao
Konda, Ravi Kumar
Challa, B. R.

Source

Journal of Analytical Methods in Chemistry

Issue

Vol. 2012, Issue 2012 (31 Dec. 2012), pp.1-8, 8 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2012-03-22

Country of Publication

Egypt

No. of Pages

8

Main Subjects

Chemistry

Abstract EN

A simple, sensitive, and rapid HPLC-MS/MS method was developed and validated for quantitative estimation of memantine in human plasma.

Chromatography was performed on Zorbax SB-C18 (4.6×75 mm, 3.5 μm) column.

Memantine (ME) and internal standard Memantine-d6(MED6) were extracted by using liquid-liquid extraction and analyzed by LC-ESI-MS/MS using multiple-reaction monitoring (MRM) mode.

The assay exhibited a linear dynamic range of 50.00–50000.00 pg/ml for ME in human plasma.

This method demonstrated an intra- and interday precision within the range of 2.1–3.7 and 1.4–7.8%, respectively.

Further intra- and interday accuracy was within the range of 95.6–99.8 and 95.7–99.1% correspondingly.

The mean recovery of ME and MED6 was 86.07±6.87 and 80.31±5.70%, respectively.

The described method was successfully employed in bioequivalence study of ME in Indian male healthy human volunteers under fasting conditions.

American Psychological Association (APA)

Konda, Ravi Kumar& Challa, B. R.& Chandu, Babu Rao& Chandrasekhar, Kothapalli B.. 2012. Bioanalytical Method Development and Validation of Memantine in Human Plasma by High Performance Liquid Chromatography with Tandem Mass Spectrometry : Application to Bioequivalence Study. Journal of Analytical Methods in Chemistry،Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-446361

Modern Language Association (MLA)

Konda, Ravi Kumar…[et al.]. Bioanalytical Method Development and Validation of Memantine in Human Plasma by High Performance Liquid Chromatography with Tandem Mass Spectrometry : Application to Bioequivalence Study. Journal of Analytical Methods in Chemistry No. 2012 (2012), pp.1-8.
https://search.emarefa.net/detail/BIM-446361

American Medical Association (AMA)

Konda, Ravi Kumar& Challa, B. R.& Chandu, Babu Rao& Chandrasekhar, Kothapalli B.. Bioanalytical Method Development and Validation of Memantine in Human Plasma by High Performance Liquid Chromatography with Tandem Mass Spectrometry : Application to Bioequivalence Study. Journal of Analytical Methods in Chemistry. 2012. Vol. 2012, no. 2012, pp.1-8.
https://search.emarefa.net/detail/BIM-446361

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-446361