A Novel Method (CiMON)‎ for Continuous Intra-Abdominal Pressure Monitoring : Pilot Test in a Pig Model

Joint Authors

Wilmer, Alexander
Dieudonne, Anne-Sophie
Wauters, Joost
Spincemaille, Liesbeth
Malbrain, Manu
Van Zwam, Kenny

Source

Critical Care Research and Practice

Issue

Vol. 2012, Issue 2012 (31 Dec. 2012), pp.1-7, 7 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2012-02-20

Country of Publication

Egypt

No. of Pages

7

Main Subjects

Medicine

Abstract EN

Background.

Intravesical pressure (IAPivp) measurement is considered to be the gold standard for assessment of intra-abdominal pressure (IAP).

This study evaluated a new minimally invasive IAP monitoring device (CiMON) against three other devices in a wide range of clinically relevant IAP and in different body positions in healthy pigs.

Methods.

The CiMON catheter (IAPCiM) and another balloon-tipped catheter (IAPspie) were positioned into the stomach.

Fluid-filled catheters were used for direct intraperitoneal (IAPdir) and IAPivp measurement.

Both in supine and 25∘ head-of-bed positions, IAP was increased from baseline to 30 mmHg.

At every IAP level, 4 IAP measurements were recorded simultaneously.

Mean differences and the limits of agreement were calculated.

Results.

Bias between IAPCiM and IAPspie was nearly zero with very good agreement, both in supine and 25∘ position.

In supine position, IAPCiM slightly overestimated IAPivp and IAPdir by 1.5 and 2.1 mmHg with reasonable agreement.

In 25∘ position, IAPCiM underestimated IAPivp and IAPdir by 1.0 and 0.5 mmHg, again with reasonable agreement.

Conclusions.

Agreement between IAPCiM and IAPspie was very good, while good-to-moderate agreement exists between IAPCiM and IAPdir or IAPivp.

Simplicity, continuous monitoring, and the combination with a feeding tube should lead to further clinical studies, evaluating this new CiMON device.

American Psychological Association (APA)

Wauters, Joost& Spincemaille, Liesbeth& Dieudonne, Anne-Sophie& Van Zwam, Kenny& Wilmer, Alexander& Malbrain, Manu. 2012. A Novel Method (CiMON) for Continuous Intra-Abdominal Pressure Monitoring : Pilot Test in a Pig Model. Critical Care Research and Practice،Vol. 2012, no. 2012, pp.1-7.
https://search.emarefa.net/detail/BIM-452430

Modern Language Association (MLA)

Wauters, Joost…[et al.]. A Novel Method (CiMON) for Continuous Intra-Abdominal Pressure Monitoring : Pilot Test in a Pig Model. Critical Care Research and Practice No. 2012 (2012), pp.1-7.
https://search.emarefa.net/detail/BIM-452430

American Medical Association (AMA)

Wauters, Joost& Spincemaille, Liesbeth& Dieudonne, Anne-Sophie& Van Zwam, Kenny& Wilmer, Alexander& Malbrain, Manu. A Novel Method (CiMON) for Continuous Intra-Abdominal Pressure Monitoring : Pilot Test in a Pig Model. Critical Care Research and Practice. 2012. Vol. 2012, no. 2012, pp.1-7.
https://search.emarefa.net/detail/BIM-452430

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-452430