Late Removal of Titanium Hardware from the Elbow Is Problematic

Joint Authors

Bachoura, Abdo
Yoshida, Ruriko
Lattermann, Christian
Kamineni, Srinath

Source

ISRN Orthopedics

Issue

Vol. 2012, Issue 2012 (31 Dec. 2012), pp.1-4, 4 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2012-02-06

Country of Publication

Egypt

No. of Pages

4

Main Subjects

Diseases
Medicine

Abstract EN

A retrospective review of 21 patients that underwent bone screw removal from the elbow was studied in relation to the type of metal, duration of implantation, and the location of the screws about the elbow.

Screw failure during extraction was the dependent variable.

Five of 21 patients experienced hardware failure during extraction.

Fourteen patients had titanium alloy implants.

In four cases, titanium screws broke during extraction.

Compared to stainless steel, titanium screw failure during removal was not statistically significant (P=0.61).

Screw removal 12 months after surgery was more likely to result in broken, retained screws in general (P=0.046) and specifically for titanium alloy (P=0.003).

Bone screws removed from the distal humerus or proximal ulna had an equal chance of fracturing (P=0.28).

There appears to be a time-related association of titanium alloy bone screw failure during hardware removal cases from the elbow.

This may be explained by titanium’s properties and osseointegration.

American Psychological Association (APA)

Bachoura, Abdo& Yoshida, Ruriko& Lattermann, Christian& Kamineni, Srinath. 2012. Late Removal of Titanium Hardware from the Elbow Is Problematic. ISRN Orthopedics،Vol. 2012, no. 2012, pp.1-4.
https://search.emarefa.net/detail/BIM-457846

Modern Language Association (MLA)

Bachoura, Abdo…[et al.]. Late Removal of Titanium Hardware from the Elbow Is Problematic. ISRN Orthopedics No. 2012 (2012), pp.1-4.
https://search.emarefa.net/detail/BIM-457846

American Medical Association (AMA)

Bachoura, Abdo& Yoshida, Ruriko& Lattermann, Christian& Kamineni, Srinath. Late Removal of Titanium Hardware from the Elbow Is Problematic. ISRN Orthopedics. 2012. Vol. 2012, no. 2012, pp.1-4.
https://search.emarefa.net/detail/BIM-457846

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-457846