Clinical and Radiographic Outcomes with a Hydroxyapatite and Porous Coated Cup Design
Joint Authors
Wang, John
Bostrom, Mathias
Padgett, Douglas
Nestor, Bryan
DiPietro, James
Westrich, Geoffrey
Source
Advances in Orthopedic Surgery
Issue
Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-5, 5 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2014-02-27
Country of Publication
Egypt
No. of Pages
5
Main Subjects
Abstract EN
Press-fit, hydroxyapatite-coated acetabular cup designs may offer a lower incidence of loosening and migration than older designs.
Our study evaluated the initial clinical and radiographic success of a cementless acetabular shell in a large cohort of primary total hip arthroplasty (THA) patients.
We queried our institution’s prospectively collected registry for a series of 771 primary THAs (695 patients) implanted with this cup by 4 high-volume arthroplasty surgeons.
Of the 613 hips with minimum 2-year followup, average HHS (Harris Hip Score) was 93.6, WOMAC (Western Ontario and McMaster Universities Osteoarthritis Index) was 87.6, and VAS (Visual Analog Scale) pain score was 1.2.
While there was a 2% reoperation rate (12 hips), none of the cups were revised for aseptic loosening.
No radiolucencies were found and there was no evidence of acetabular loosening.
At early followup, this newer cementless acetabular cup implant design exhibits high survivorship and clinical success.
American Psychological Association (APA)
Wang, John& DiPietro, James& Bostrom, Mathias& Nestor, Bryan& Padgett, Douglas& Westrich, Geoffrey. 2014. Clinical and Radiographic Outcomes with a Hydroxyapatite and Porous Coated Cup Design. Advances in Orthopedic Surgery،Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-461753
Modern Language Association (MLA)
Wang, John…[et al.]. Clinical and Radiographic Outcomes with a Hydroxyapatite and Porous Coated Cup Design. Advances in Orthopedic Surgery No. 2014 (2014), pp.1-5.
https://search.emarefa.net/detail/BIM-461753
American Medical Association (AMA)
Wang, John& DiPietro, James& Bostrom, Mathias& Nestor, Bryan& Padgett, Douglas& Westrich, Geoffrey. Clinical and Radiographic Outcomes with a Hydroxyapatite and Porous Coated Cup Design. Advances in Orthopedic Surgery. 2014. Vol. 2014, no. 2014, pp.1-5.
https://search.emarefa.net/detail/BIM-461753
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-461753