Improvements on Noninvasive Blood Glucose Biosensors Using Wavelets for Quick Fault Detection

Joint Authors

Campetelli, Germán
Basualdo, Marta
Zumoffen, David

Source

Journal of Sensors

Issue

Vol. 2011, Issue 2011 (31 Dec. 2011), pp.1-11, 11 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2011-06-02

Country of Publication

Egypt

No. of Pages

11

Main Subjects

Civil Engineering

Abstract EN

Noninvasive blood glucose sensors are still under development stage considering that they are far from being suitable for use in anartificial pancreas.

The latter has three main parts: the blood glucose sensor, the insulin pump and the controller.

However, for the biosensor analyzed here, some common failures such as signal shifts and unreal picks were found.

They must be taken into account, for computing the correct insulin dosage for diabetic persons.

Hence, a fault detection system based on discrete wavelets transform (DWT) is applied here.

The main idea is, when the fault occurs, to do a proper measurement compensation for sending the corrected value to the predictive functional controller (PFC) algorithm.

The study is done by reproducing the fault on the blood glucose measurements.

They are obtained from a mathematical model of the endocrine system of an adult diabetic patient.

This model was approved by the FDA in 2008.

Then, the simulation environment includes faulty blood glucose measurements and a fault diagnosis and identification (FDI) system based on DWT.

The FDI system gives to the PFC algorithm the correct information to turn it into a fault-tolerant controller (FTC).

The main goal is to deliver the correct insulin dosage to the patient.

American Psychological Association (APA)

Campetelli, Germán& Zumoffen, David& Basualdo, Marta. 2011. Improvements on Noninvasive Blood Glucose Biosensors Using Wavelets for Quick Fault Detection. Journal of Sensors،Vol. 2011, no. 2011, pp.1-11.
https://search.emarefa.net/detail/BIM-466409

Modern Language Association (MLA)

Campetelli, Germán…[et al.]. Improvements on Noninvasive Blood Glucose Biosensors Using Wavelets for Quick Fault Detection. Journal of Sensors No. 2011 (2011), pp.1-11.
https://search.emarefa.net/detail/BIM-466409

American Medical Association (AMA)

Campetelli, Germán& Zumoffen, David& Basualdo, Marta. Improvements on Noninvasive Blood Glucose Biosensors Using Wavelets for Quick Fault Detection. Journal of Sensors. 2011. Vol. 2011, no. 2011, pp.1-11.
https://search.emarefa.net/detail/BIM-466409

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-466409