Intracochlear Fluid Pressure Changes Related to the Insertional Speed of a CI Electrode

Joint Authors

Todt, Ingo
Ernst, Arne
Mittmann, P.

Source

BioMed Research International

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-4, 4 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-07-16

Country of Publication

Egypt

No. of Pages

4

Main Subjects

Medicine

Abstract EN

Introduction.

To preserve residual hearing the atraumaticity of the cochlea electrode insertion has become a focus of cochlear implant research.

In addition to other factors, the speed of insertion is thought to be a contributing factor in the concept of atraumatic implantation.

The aim of our study was to observe intracochlear fluid pressure changes due to different insertional speeds of an implant electrode in a cochlear model.

Materials and Methods.

The experiments were performed using an artificial cochlear model.

A linear actuator was mounted on an Advanced Bionics IJ insertional tool.

The intracochlear fluid pressure was recorded through a pressure sensor which was placed in the helicotrema area.

Defined insertions were randomly performed with speeds of 0.1 mm/sec, 0.25 mm/sec, 0.5 mm/sec, 1 mm/sec, and 2 mm/sec.

Results.

A direct correlation between speed and pressure was observed.

Mean maximum values of intracochlear fluid pressure varied between 0.41 mm Hg and 1.27 mm Hg.

Conclusion.

We provide the first results of fluid pressure changes due to insertional speeds of CI electrodes in a cochlear model.

A relationship between the insertional speed and intracochlear fluid pressure was observed.

Further experiments are needed to apply these results to the in vivo situation.

American Psychological Association (APA)

Todt, Ingo& Mittmann, P.& Ernst, Arne. 2014. Intracochlear Fluid Pressure Changes Related to the Insertional Speed of a CI Electrode. BioMed Research International،Vol. 2014, no. 2014, pp.1-4.
https://search.emarefa.net/detail/BIM-477123

Modern Language Association (MLA)

Todt, Ingo…[et al.]. Intracochlear Fluid Pressure Changes Related to the Insertional Speed of a CI Electrode. BioMed Research International No. 2014 (2014), pp.1-4.
https://search.emarefa.net/detail/BIM-477123

American Medical Association (AMA)

Todt, Ingo& Mittmann, P.& Ernst, Arne. Intracochlear Fluid Pressure Changes Related to the Insertional Speed of a CI Electrode. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-4.
https://search.emarefa.net/detail/BIM-477123

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-477123