Does Vertebroplasty Affect Radiation Dose Distribution? : Comparison of Spatial Dose Distributions in a Cement-Injected Vertebra as Calculated by Treatment Planning System and Actual Spatial Dose Distribution

Joint Authors

Sawada, Satoshi
Tanigawa, Noboru
Yagi, Rie
Komemushi, Atsushi
Kariya, Shuji
Ikeda, Koshi
Utsunomiya, Keita
Harima, Yoko
Suzuki, Satoshi
Nakatani, Miyuki
Sano, Akira

Source

Radiology Research and Practice

Issue

Vol. 2012, Issue 2012 (31 Dec. 2012), pp.1-6, 6 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2012-04-10

Country of Publication

Egypt

No. of Pages

6

Main Subjects

Medicine

Abstract EN

Purpose.

To assess differences in dose distribution of a vertebral body injected with bone cement as calculated by radiation treatment planning system (RTPS) and actual dose distribution.

Methods.

We prepared two water-equivalent phantoms with cement, and the other two phantoms without cement.

The bulk density of the bone cement was imported into RTPS to reduce error from high CT values.

A dose distribution map for the phantoms with and without cement was calculated using RTPS with clinical setting and with the bulk density importing.

Actual dose distribution was measured by the film density.

Dose distribution as calculated by RTPS was compared to the dose distribution measured by the film dosimetry.

Results.

For the phantom with cement, dose distribution was distorted for the areas corresponding to inside the cement and on the ventral side of the cement.

However, dose distribution based on film dosimetry was undistorted behind the cement and dose increases were seen inside cement and around the cement.

With the equivalent phantom with bone cement, differences were seen between dose distribution calculated by RTPS and that measured by the film dosimetry.

Conclusion.

The dose distribution of an area containing bone cement calculated using RTPS differs from actual dose distribution.

American Psychological Association (APA)

Komemushi, Atsushi& Tanigawa, Noboru& Kariya, Shuji& Yagi, Rie& Nakatani, Miyuki& Suzuki, Satoshi…[et al.]. 2012. Does Vertebroplasty Affect Radiation Dose Distribution? : Comparison of Spatial Dose Distributions in a Cement-Injected Vertebra as Calculated by Treatment Planning System and Actual Spatial Dose Distribution. Radiology Research and Practice،Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-481678

Modern Language Association (MLA)

Komemushi, Atsushi…[et al.]. Does Vertebroplasty Affect Radiation Dose Distribution? : Comparison of Spatial Dose Distributions in a Cement-Injected Vertebra as Calculated by Treatment Planning System and Actual Spatial Dose Distribution. Radiology Research and Practice No. 2012 (2012), pp.1-6.
https://search.emarefa.net/detail/BIM-481678

American Medical Association (AMA)

Komemushi, Atsushi& Tanigawa, Noboru& Kariya, Shuji& Yagi, Rie& Nakatani, Miyuki& Suzuki, Satoshi…[et al.]. Does Vertebroplasty Affect Radiation Dose Distribution? : Comparison of Spatial Dose Distributions in a Cement-Injected Vertebra as Calculated by Treatment Planning System and Actual Spatial Dose Distribution. Radiology Research and Practice. 2012. Vol. 2012, no. 2012, pp.1-6.
https://search.emarefa.net/detail/BIM-481678

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-481678