Dark Signal Temperature Dependence Correction Method for Miniature Spectrometer Modules
Author
Source
Issue
Vol. 2011, Issue 2011 (31 Dec. 2011), pp.1-9, 9 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2011-04-05
Country of Publication
Egypt
No. of Pages
9
Main Subjects
Abstract EN
A dark signal temperature dependence correction method for miniature spectrometer modules is described in this paper.
It is based on laboratory measurements of dark signal temperature dependence at few different integration times.
A set of parameters are calculated which make it possible to estimate dark signal at any temperature and integration time within reasonable range.
In field conditions, it is not always possible to take frequent dark signal readings during spectral measurements.
If temperature is recorded during the measurement, this method can be used for estimating dark signal for every single spectral measurement.
The method is validated on two different miniature spectrometers.
American Psychological Association (APA)
Kuusk, Joel. 2011. Dark Signal Temperature Dependence Correction Method for Miniature Spectrometer Modules. Journal of Sensors،Vol. 2011, no. 2011, pp.1-9.
https://search.emarefa.net/detail/BIM-484715
Modern Language Association (MLA)
Kuusk, Joel. Dark Signal Temperature Dependence Correction Method for Miniature Spectrometer Modules. Journal of Sensors No. 2011 (2011), pp.1-9.
https://search.emarefa.net/detail/BIM-484715
American Medical Association (AMA)
Kuusk, Joel. Dark Signal Temperature Dependence Correction Method for Miniature Spectrometer Modules. Journal of Sensors. 2011. Vol. 2011, no. 2011, pp.1-9.
https://search.emarefa.net/detail/BIM-484715
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-484715