Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD

Joint Authors

Broers, Bernard
van Opstal, Jurren
van Gelder, Berry M.
Debie, Luuk

Source

Case Reports in Cardiology

Issue

Vol. 2011, Issue 2011 (31 Dec. 2011), pp.1-3, 3 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2011-10-15

Country of Publication

Egypt

No. of Pages

3

Main Subjects

Diseases

Abstract EN

We report atrio-ventricular cross talk in a patient with a dual chamber ICD leading to ventricular pacing inhibition.

This was observed in an ICD without the ventricular safety pace option, which normally is a sufficient protection against this phenomenon.

Cross talk could be prevented effectively by reprogramming ventricular sensitivity to a less sensitive setting.

American Psychological Association (APA)

Debie, Luuk& Broers, Bernard& van Opstal, Jurren& van Gelder, Berry M.. 2011. Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD. Case Reports in Cardiology،Vol. 2011, no. 2011, pp.1-3.
https://search.emarefa.net/detail/BIM-486043

Modern Language Association (MLA)

Debie, Luuk…[et al.]. Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD. Case Reports in Cardiology No. 2011 (2011), pp.1-3.
https://search.emarefa.net/detail/BIM-486043

American Medical Association (AMA)

Debie, Luuk& Broers, Bernard& van Opstal, Jurren& van Gelder, Berry M.. Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD. Case Reports in Cardiology. 2011. Vol. 2011, no. 2011, pp.1-3.
https://search.emarefa.net/detail/BIM-486043