Abstract EN
We report atrio-ventricular cross talk in a patient with a dual chamber ICD leading to ventricular pacing inhibition.
This was observed in an ICD without the ventricular safety pace option, which normally is a sufficient protection against this phenomenon.
Cross talk could be prevented effectively by reprogramming ventricular sensitivity to a less sensitive setting.
American Psychological Association (APA)
Debie, Luuk& Broers, Bernard& van Opstal, Jurren& van Gelder, Berry M.. 2011. Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD. Case Reports in Cardiology،Vol. 2011, no. 2011, pp.1-3.
https://search.emarefa.net/detail/BIM-486043
Modern Language Association (MLA)
Debie, Luuk…[et al.]. Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD. Case Reports in Cardiology No. 2011 (2011), pp.1-3.
https://search.emarefa.net/detail/BIM-486043
American Medical Association (AMA)
Debie, Luuk& Broers, Bernard& van Opstal, Jurren& van Gelder, Berry M.. Atrioventricular Cross-Talk Leading to Ventricular Pacing Inhibition in a Dual-Chamber ICD. Case Reports in Cardiology. 2011. Vol. 2011, no. 2011, pp.1-3.
https://search.emarefa.net/detail/BIM-486043
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-486043