An Optimized Design of Single-Channel Beta-Gamma Coincidence Phoswich Detector by Geant4 Monte Carlo Simulations

Joint Authors

Zhang, Weihua
Ungar, Kurt
Yi, Jing
Bean, Marc
Mekarski, Pawel

Source

Science and Technology of Nuclear Installations

Issue

Vol. 2011, Issue 2011 (31 Dec. 2011), pp.1-5, 5 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2011-05-14

Country of Publication

Egypt

No. of Pages

5

Main Subjects

Engineering Sciences and Information Technology

Abstract EN

An optimized single-channel phoswich well detector design has been proposed and assessed in order to improve beta-gamma coincidence measurement sensitivity of xenon radioisotopes.

This newly designed phoswich well detector consists of a plastic beta counting cell (BC404) embedded in a CsI(Tl) crystal coupled to a photomultiplier tube.

The BC404 is configured in a cylindrical pipe shape to minimise light collection deterioration.

The CsI(Tl) crystal consists of a rectangular part and a semicylindrical scintillation part as a light reflector to increase light gathering.

Compared with a PhosWatch detector, the final optimized detector geometry showed 15% improvement in the energy resolution of a 131mXe 129.4 keV conversion electron peak.

The predicted beta-gamma coincidence efficiencies of xenon radioisotopes have also been improved accordingly.

American Psychological Association (APA)

Zhang, Weihua& Mekarski, Pawel& Bean, Marc& Yi, Jing& Ungar, Kurt. 2011. An Optimized Design of Single-Channel Beta-Gamma Coincidence Phoswich Detector by Geant4 Monte Carlo Simulations. Science and Technology of Nuclear Installations،Vol. 2011, no. 2011, pp.1-5.
https://search.emarefa.net/detail/BIM-495036

Modern Language Association (MLA)

Zhang, Weihua…[et al.]. An Optimized Design of Single-Channel Beta-Gamma Coincidence Phoswich Detector by Geant4 Monte Carlo Simulations. Science and Technology of Nuclear Installations No. 2011 (2011), pp.1-5.
https://search.emarefa.net/detail/BIM-495036

American Medical Association (AMA)

Zhang, Weihua& Mekarski, Pawel& Bean, Marc& Yi, Jing& Ungar, Kurt. An Optimized Design of Single-Channel Beta-Gamma Coincidence Phoswich Detector by Geant4 Monte Carlo Simulations. Science and Technology of Nuclear Installations. 2011. Vol. 2011, no. 2011, pp.1-5.
https://search.emarefa.net/detail/BIM-495036

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-495036