Intraocular Microsurgical Forceps (20, 23, and 25 gauge)‎ Membrane Peeling Forces Assessment

Joint Authors

Olson, Jeffrey L.
Quiroz-Mercado, Hugo
Oliver, Scott C. N.
Patel, Chirag
Mandava, Naresh
Velez-Montoya, Raul

Source

Journal of Ophthalmology

Issue

Vol. 2013, Issue 2013 (31 Dec. 2013), pp.1-4, 4 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2013-06-24

Country of Publication

Egypt

No. of Pages

4

Main Subjects

Medicine

Abstract EN

Background.

To assess the peeling forces exerted by different calibers of microsurgical forceps on an experimental model of epiretinal membrane.

Methods.

A model of epiretinal membrane was constructed using thin cellulose paper and heptanes-isopropyl alcohol 1% mixture.

The model was mounted on a force censoring device.

Subsequently, flaps were created with three different microsurgical forceps of different calibers.

We recorded the number of attempts, the duration of the event, and the pushing and the pulling forces during the peeling.

The results were compared by a one-way ANOVA and a Fisher unprotected least significant difference test with an alpha value of 0.05 for statistically significance.

Results.

There was a statistical significant difference on the pulling and pushing forces between the 25 gauge (13.79 mN; −13.27 mN) and the 23 (6.63 mN; −5.76 mN) and 20 (5.02 mN; −5.30 mN) gauge, being greater in the first (P<0.001).

There were no differences in the duration of all events, meaning that all the forces were measured within the same period of time.

Conclusions.

The 25 gauge microsurgical forceps exerted the greatest mechanical stress over our simulated epiretinal membrane model and required more attempts to create a surgical suitable flap.

The clinical implication of this finding is still to be determined.

American Psychological Association (APA)

Velez-Montoya, Raul& Patel, Chirag& Oliver, Scott C. N.& Quiroz-Mercado, Hugo& Mandava, Naresh& Olson, Jeffrey L.. 2013. Intraocular Microsurgical Forceps (20, 23, and 25 gauge) Membrane Peeling Forces Assessment. Journal of Ophthalmology،Vol. 2013, no. 2013, pp.1-4.
https://search.emarefa.net/detail/BIM-497801

Modern Language Association (MLA)

Velez-Montoya, Raul…[et al.]. Intraocular Microsurgical Forceps (20, 23, and 25 gauge) Membrane Peeling Forces Assessment. Journal of Ophthalmology No. 2013 (2013), pp.1-4.
https://search.emarefa.net/detail/BIM-497801

American Medical Association (AMA)

Velez-Montoya, Raul& Patel, Chirag& Oliver, Scott C. N.& Quiroz-Mercado, Hugo& Mandava, Naresh& Olson, Jeffrey L.. Intraocular Microsurgical Forceps (20, 23, and 25 gauge) Membrane Peeling Forces Assessment. Journal of Ophthalmology. 2013. Vol. 2013, no. 2013, pp.1-4.
https://search.emarefa.net/detail/BIM-497801

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID