Achieving Accurate Ligament Balancing Using Robotic-Assisted Unicompartmental Knee Arthroplasty
Joint Authors
Lang, Jason E.
Mofidi, Ali
Mannava, Sandeep
Conditt, Michael A.
Plate, Johannes F.
Poehling, Gary G.
Smith, Beth P.
Jinnah, Riyaz H.
Source
Issue
Vol. 2013, Issue 2013 (31 Dec. 2013), pp.1-6, 6 p.
Publisher
Hindawi Publishing Corporation
Publication Date
2013-03-24
Country of Publication
Egypt
No. of Pages
6
Main Subjects
Abstract EN
Unicompartmental knee arthroplasty (UKA) allows replacement of a single compartment in patients with limited disease.
However, UKA is technically challenging and relies on accurate component positioning and restoration of natural knee kinematics.
This study examined the accuracy of dynamic, real-time ligament balancing using a robotic-assisted UKA system.
Surgical data obtained from the computer system were prospectively collected from 51 patients (52 knees) undergoing robotic-assisted medial UKA by a single surgeon.
Dynamic ligament balancing of the knee was obtained under valgus stress prior to component implantation and then compared to final ligament balance with the components in place.
Ligament balancing was accurate up to 0.53 mm compared to the preoperative plan, with 83% of cases within 1 mm at 0°, 30°, 60°, 90°, and 110° of flexion.
Ligamentous laxity of 1.31±0.13 mm at 30° of flexion was corrected successfully to 0.78±0.17 mm (P<0.05).
Robotic-assisted UKA allows accurate and precise reproduction of a surgical balance plan using dynamic, real-time soft-tissue balancing to help restore natural knee kinematics, potentially improving implant survival and functional outcomes.
American Psychological Association (APA)
Plate, Johannes F.& Mofidi, Ali& Mannava, Sandeep& Smith, Beth P.& Lang, Jason E.& Poehling, Gary G.…[et al.]. 2013. Achieving Accurate Ligament Balancing Using Robotic-Assisted Unicompartmental Knee Arthroplasty. Advances in Orthopedics،Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-502079
Modern Language Association (MLA)
Plate, Johannes F.…[et al.]. Achieving Accurate Ligament Balancing Using Robotic-Assisted Unicompartmental Knee Arthroplasty. Advances in Orthopedics No. 2013 (2013), pp.1-6.
https://search.emarefa.net/detail/BIM-502079
American Medical Association (AMA)
Plate, Johannes F.& Mofidi, Ali& Mannava, Sandeep& Smith, Beth P.& Lang, Jason E.& Poehling, Gary G.…[et al.]. Achieving Accurate Ligament Balancing Using Robotic-Assisted Unicompartmental Knee Arthroplasty. Advances in Orthopedics. 2013. Vol. 2013, no. 2013, pp.1-6.
https://search.emarefa.net/detail/BIM-502079
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references
Record ID
BIM-502079