A Novel Pedicle Screw with Mobile Connection : A Pilot Study

Joint Authors

Tokuhashi, Yasuaki
Ajiro, Yasumitsu
Uei, Hiroshi
Oshima, Masashi

Source

BioMed Research International

Issue

Vol. 2014, Issue 2014 (31 Dec. 2014), pp.1-6, 6 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2014-03-03

Country of Publication

Egypt

No. of Pages

6

Main Subjects

Medicine

Abstract EN

To prevent adjacent disc problems after spinal fusion, a pedicle screw with a mobile junction between the head and threaded shaft was newly developed.

The threaded shaft of the screw has 10 degrees mobility in all directions, but its structure is to prevent abnormal translation and tilting.

This screw was evaluated as follows: (1) endurance test: 106 times rotational stress was applied; (2) biological reactions: novel screws with a mobile head and conventional screws with a fixed head were inserted into the bilateral pedicles of the L3, L4, and L5 in two mini pigs with combination.

Eight months after surgery, vertebral units with the screw rod constructs were collected.

After CT scan, the soft and bony tissues around the screws were examined grossly and histologically.

As a result, none of the screws broke during the endurance test stressing.

The mean amount of abrasion wear was 0.0338 g.

In the resected mini pig section, though zygapophyseal joints between fixed-head screws showed bony union, the amount of callus in the zygapophyseal joints connected with mobile-head screws was small, and joint space was confirmed by CT.

No metalloses were noted around any of the screws.

Novel screws were suggested to be highly durable and histologically safe.

American Psychological Association (APA)

Tokuhashi, Yasuaki& Oshima, Masashi& Ajiro, Yasumitsu& Uei, Hiroshi. 2014. A Novel Pedicle Screw with Mobile Connection : A Pilot Study. BioMed Research International،Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-502553

Modern Language Association (MLA)

Tokuhashi, Yasuaki…[et al.]. A Novel Pedicle Screw with Mobile Connection : A Pilot Study. BioMed Research International No. 2014 (2014), pp.1-6.
https://search.emarefa.net/detail/BIM-502553

American Medical Association (AMA)

Tokuhashi, Yasuaki& Oshima, Masashi& Ajiro, Yasumitsu& Uei, Hiroshi. A Novel Pedicle Screw with Mobile Connection : A Pilot Study. BioMed Research International. 2014. Vol. 2014, no. 2014, pp.1-6.
https://search.emarefa.net/detail/BIM-502553

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID