Embedded Spiral Microstrip Implantable Antenna

Joint Authors

Kishk, Ahmed A.
Huang, Wei

Source

International Journal of Antennas and Propagation

Issue

Vol. 2011, Issue 2011 (31 Dec. 2011), pp.1-6, 6 p.

Publisher

Hindawi Publishing Corporation

Publication Date

2011-08-18

Country of Publication

Egypt

No. of Pages

6

Main Subjects

Electronic engineering

Abstract EN

A miniaturized embedded spiral-shaped microstrip antenna simulated within a block of 2/3 human muscle phantom is designed for implanted cardiac pacemaker.

The new design not only achieves the good matching at the required 403 MHz but also obtains 57% size reduction over the existing design.

The reflection coefficients are measured in both free space and in the human-tissue-equivalent dielectric liquid.

Good agreement is achieved between simulated results and the experimental results.

American Psychological Association (APA)

Huang, Wei& Kishk, Ahmed A.. 2011. Embedded Spiral Microstrip Implantable Antenna. International Journal of Antennas and Propagation،Vol. 2011, no. 2011, pp.1-6.
https://search.emarefa.net/detail/BIM-508171

Modern Language Association (MLA)

Huang, Wei& Kishk, Ahmed A.. Embedded Spiral Microstrip Implantable Antenna. International Journal of Antennas and Propagation No. 2011 (2011), pp.1-6.
https://search.emarefa.net/detail/BIM-508171

American Medical Association (AMA)

Huang, Wei& Kishk, Ahmed A.. Embedded Spiral Microstrip Implantable Antenna. International Journal of Antennas and Propagation. 2011. Vol. 2011, no. 2011, pp.1-6.
https://search.emarefa.net/detail/BIM-508171

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references

Record ID

BIM-508171