Accuracy of platelet Ccounting by optical and impedance methods in patientsthrombocytopaenia and microcytosis
Other Title(s)
الدقة في عد الصفيحات بواسطة طرق بصرية و معاوقية عند مرضى مصابين بقلة الصفيحات و صغر الكريات الحمراء
Joint Authors
al-Farisi, Rayya
al-Hashimi, Sulayman
al-Riyami, Hamad
Khan, Hamad
al-Kindi, Salam
Bu lassil, Muhammad Rashid
Source
Sultan Qaboos University Medical Journal
Issue
Vol. 15, Issue 4 (30 Nov. 2015), pp.463-468, 6 p.
Publisher
Sultan Qaboos University College of Medicine and Health Sciences
Publication Date
2015-11-30
Country of Publication
Oman
No. of Pages
6
Main Subjects
Topics
Abstract AR
الهدف : يمثل الحصول على تعداد دقيق للصفيحات في عينات دم المصابين بصغر الكريات الحمراء تحديا كبيرا حتى عند استخدام أكثر أجهزة التحليل الأتوماتيكية ورثوقية.
و يقوم جهاز سيل داين CELL-DYNTM (من معامل أبوت بشيكاغو- ولاية الينوي-الولايات المتحدة) باستخدام الكثافة البصرية و المعاوقية الكهربائية لعد الصفيحات.
و يهدف هذا البحث لتقويم دقة الطرق البصرية و المعاوقية الكهربائية لتحديد أعداد الصفيحات بدقة في عينات دماء المرضى المصابين بقلة الصفيحات و صغر الكريات الحمراء، و المعرفة بصغر حجم الكرية الوسطى (MCV).
و تم كذلك بحث تأثير صغر الكريات الحمراء على دقة عد الصفيحات.
الطريقة : أجريت الدراسة بين فبراير و ديسمبر2014م.
بمختبر علم الدم بمستشفى جامعة السلطان قابوس بمسقط في عمان.
و جمع عينات دم من 189 من المرضى المصابين بقلة الصفيحات و قيم MCV أقل من 76 > فيمتوليتر.
و تم عد الصفيحات عن طريق استخدام الكثافة البصرية و المعاوقية الكهربائية.
و تم فحص شرائح دم محيطي من كل عينة دم كمرجعية لعد الصفيحات.
النتائج : كانت نتائج عد الصفيحات بواسطة الطريقة المعاوقية تفوق تلك المتحصل عليها من بالطريقة البصرية بنحو % 30 في المتوسط (0.001 > P) و كان معامل الارتباط هو 0.52 (% 95 فاصل الموثوقية : 0.62-0.41) مما يؤشر إلى اعتمادية متوسطة عند الطريقتين.
و تراوحت درجة التوافق بين الطريقتين بين 85.5- و 24.3، مع تحيز (خطأ منهجي) يبلغ 30-.
مما يدل على أن الطريقتين تخرجان بنتائج مختلفة لعدد الصفيحات.
الخلاصة : النتائج المتحصل عليها من الطريقة المعاوقية هي أكبر من غيرها عند استخدامها في دماء مرضى مصابين بقلة الصفيحات و صغر الكريات الحمراء.
لذا ينبغي الاهتمام أكثر بتحسين الدقة، خاصة عند المرضى المصابين بصغر الكريات الحمراء.
Abstract EN
Obtaining accurate platelet counts in microcytic blood samples is challenging, even with the most reliable automated haematology analysers.
The CELL-DYN™ Sapphire (Abbott Laboratories, Chicago, Illinois, USA) analyser uses both optical density and electronic impedance methods for platelet counting.
This study aimed to evaluate the accuracy of optical density and electrical impedance methods in determining true platelet counts in thrombocytopaenic samples with microcytosis as defined by low mean corpuscular volume (MCV) of red blood cells.
Additionally, the impact of microcytosis on platelet count accuracy was evaluated.
Methods: This study was carried out between February and December 2014 at the Haematology Laboratory of the Sultan Qaboos University Hospital in Muscat, Oman.
Blood samples were collected and analysed from 189 patients with thrombocytopaenia and MCV values of <76 femtolitres.
Platelet counts were tested using both optical and impedance methods.
Stained peripheral blood films for each sample were then reviewed as a reference method to confirm platelet counts.
Results: The platelet counts estimated by the impedance method were on average 30% higher than those estimated by the optical method (P <0.001).
The estimated intraclass correlation coefficient was 0.52 (95% confidence interval: 0.41–0.62), indicating moderate reliability between the methods.
The degree of agreement between methods ranged from -85.5 to 24.3 with an estimated bias of -30, suggesting that these methods generate different platelet results.
Conclusion: The impedance method significantly overestimated platelet counts in microcytic and thrombocytopaenic blood samples.
Further attention is therefore needed to improve the accuracy of platelet counts, particularly for patients with conditions associated with microcytosis.
American Psychological Association (APA)
Bu lassil, Muhammad Rashid& al-Farisi, Rayya& al-Hashimi, Sulayman& al-Riyami, Hamad& Khan, Hamad& al-Kindi, Salam. 2015. Accuracy of platelet Ccounting by optical and impedance methods in patientsthrombocytopaenia and microcytosis. Sultan Qaboos University Medical Journal،Vol. 15, no. 4, pp.463-468.
https://search.emarefa.net/detail/BIM-645471
Modern Language Association (MLA)
Bu lassil, Muhammad Rashid…[et al.]. Accuracy of platelet Ccounting by optical and impedance methods in patientsthrombocytopaenia and microcytosis. Sultan Qaboos University Medical Journal Vol. 15, no. 4 (Nov. 2015), pp.463-468.
https://search.emarefa.net/detail/BIM-645471
American Medical Association (AMA)
Bu lassil, Muhammad Rashid& al-Farisi, Rayya& al-Hashimi, Sulayman& al-Riyami, Hamad& Khan, Hamad& al-Kindi, Salam. Accuracy of platelet Ccounting by optical and impedance methods in patientsthrombocytopaenia and microcytosis. Sultan Qaboos University Medical Journal. 2015. Vol. 15, no. 4, pp.463-468.
https://search.emarefa.net/detail/BIM-645471
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references : p. 468
Record ID
BIM-645471