Validation of GEANT4 monte carlo simulation code for 6 MV varian linac photon beam

Joint Authors

Salamah, E.
Radi, A.
Imad, N.
Ali, A. S.

Source

Arab Journal of Nuclear Sciences and Applications

Issue

Vol. 48, Issue 2 (30 Apr. 2015), pp.70-76, 7 p.

Publisher

The Egyptian Society of Nuclear Science and Applications

Publication Date

2015-04-30

Country of Publication

Egypt

No. of Pages

7

Main Subjects

Physics
Medicine

Abstract EN

The head of a clinical linear accelerator based on the manufacturer’s detailed information is simulated by using GEANT4.

Percentage Depth Dose (PDD) and flatness symmetry (lateral dose profiles) in water phantom were evaluated.

Comparisons between experimental and simulated data were carried out for two field sizes; 5 × 5, and 10 × 10 cm2.

The obtained results indicated that GEANT4 code is a promising and validated Monte Carlo program for using in radiotherapy applications.

American Psychological Association (APA)

Salamah, E.& Ali, A. S.& Imad, N.& Radi, A.. 2015. Validation of GEANT4 monte carlo simulation code for 6 MV varian linac photon beam. Arab Journal of Nuclear Sciences and Applications،Vol. 48, no. 2, pp.70-76.
https://search.emarefa.net/detail/BIM-724250

Modern Language Association (MLA)

Salamah, E.…[et al.]. Validation of GEANT4 monte carlo simulation code for 6 MV varian linac photon beam. Arab Journal of Nuclear Sciences and Applications Vol. 48, no. 2 (Apr. 2015), pp.70-76.
https://search.emarefa.net/detail/BIM-724250

American Medical Association (AMA)

Salamah, E.& Ali, A. S.& Imad, N.& Radi, A.. Validation of GEANT4 monte carlo simulation code for 6 MV varian linac photon beam. Arab Journal of Nuclear Sciences and Applications. 2015. Vol. 48, no. 2, pp.70-76.
https://search.emarefa.net/detail/BIM-724250

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references : p. 76

Record ID