Calibration of cylindrical NaI (Tl)‎ detector using different perpendicular radioactive cylindrical sources

Author

Badawi, Muhammad S.

Source

Arab Journal of Nuclear Sciences and Applications

Issue

Vol. 48, Issue 4 (31 Oct. 2015), pp.67-75, 9 p.

Publisher

The Egyptian Society of Nuclear Science and Applications

Publication Date

2015-10-31

Country of Publication

Egypt

No. of Pages

9

Main Subjects

Nutrition & Dietetics

Abstract EN

Environmental samples containing some contaminated samples of a wide variety of characteristics are routinely analyzed using gamma-ray detectors.

In this work the full energy peak efficiency of NaI(Tl) scintillation detectors 2”x2” and 3”x3” inches have been calculated experimentally by using perpendicular radioactive cylindrical sources and theoretically by applying an analytical efficiency transfer expression.

This expression was based on including the ratio of the source-to-detector effective solid angle.

Measurements were made using the perpendicular cylindrical vials filled with 152Eu solution.

The Calibration process covers a wide energy range starting from 121 up to 1408 keV.

In addition, the self-attenuation factor was estimated by using the linear attenuation factor of the assayed source material and the geometrical configuration of the assay set-up at low source-detector distance.

A remarkable agreement between measured and calculated efficiency values was observed.

American Psychological Association (APA)

Badawi, Muhammad S.. 2015. Calibration of cylindrical NaI (Tl) detector using different perpendicular radioactive cylindrical sources. Arab Journal of Nuclear Sciences and Applications،Vol. 48, no. 4, pp.67-75.
https://search.emarefa.net/detail/BIM-724671

Modern Language Association (MLA)

Badawi, Muhammad S.. Calibration of cylindrical NaI (Tl) detector using different perpendicular radioactive cylindrical sources. Arab Journal of Nuclear Sciences and Applications Vol. 48, no. 4 (Oct. 2015), pp.67-75.
https://search.emarefa.net/detail/BIM-724671

American Medical Association (AMA)

Badawi, Muhammad S.. Calibration of cylindrical NaI (Tl) detector using different perpendicular radioactive cylindrical sources. Arab Journal of Nuclear Sciences and Applications. 2015. Vol. 48, no. 4, pp.67-75.
https://search.emarefa.net/detail/BIM-724671

Data Type

Journal Articles

Language

English

Notes

Includes bibliographical references : p. 75

Record ID

BIM-724671