Assessment of ISNAG fluorimeter (total fluorescence measurements at+90° &-90° using four solar cell on each side for 100mm distance at 2mm path length) with well-known fluorescent molecules via CFIA
Other Title(s)
تقييم مقياس الفلور ISNAG (القياس للفلورة الكليةعلى +09 °و -09°باستخدام أربعة خلايا شمسية على كل جانب لمسافة 100mm و بمسار تشعيع 2mm) مع جزيئات الفلورة المعروفة عن طريق تقنية الحقن الجرياني المستمر
Joint Authors
Turki, Nagam Shakir
Munshid, Haydar Qays
Source
Issue
Vol. 59, Issue 1B (31 Mar. 2018), pp.240-250, 11 p.
Publisher
University of Baghdad College of Science
Publication Date
2018-03-31
Country of Publication
Iraq
No. of Pages
11
Main Subjects
Topics
Abstract EN
Two well-known fluorescent molecules: fluorescein sodium salt (FSS) and 2,7- dichloro fluorescein (DCF) were tried to prove the efficiency, trustability and repeatability of ISNAG fluorimeter by using discrete and continuous flow injection analysis modes.A linear range of 0.002-1 mmol/L for FSS and 0.003-0.7 mmol/L was for DCF, with LOD 0.0018 mmol/L and 0.002 mmol/L for FSS and DCF respectively, were obtained for discrete mode of analysis.
While the continuous mode gave a linear range of 0.002-0.7 mmol/L and 0.003-0.5 mmol/L for FSS and DCF respectively, the LOD were 0.0016mmol/L and 0.0018 mmol/L for FSS and DCF respectively.
The results were compared with classical method at variable λex for both fluorescent molecules at 95% confidence level.
The comparison data shows that ISNAG fluorimeter is the choice with excellent extended detection and a wider applicability.
American Psychological Association (APA)
Turki, Nagam Shakir& Munshid, Haydar Qays. 2018. Assessment of ISNAG fluorimeter (total fluorescence measurements at+90° &-90° using four solar cell on each side for 100mm distance at 2mm path length) with well-known fluorescent molecules via CFIA. Iraqi Journal of Science،Vol. 59, no. 1B, pp.240-250.
https://search.emarefa.net/detail/BIM-832069
Modern Language Association (MLA)
Turki, Nagam Shakir& Munshid, Haydar Qays. Assessment of ISNAG fluorimeter (total fluorescence measurements at+90° &-90° using four solar cell on each side for 100mm distance at 2mm path length) with well-known fluorescent molecules via CFIA. Iraqi Journal of Science Vol. 59, no. 1B (2018), pp.240-250.
https://search.emarefa.net/detail/BIM-832069
American Medical Association (AMA)
Turki, Nagam Shakir& Munshid, Haydar Qays. Assessment of ISNAG fluorimeter (total fluorescence measurements at+90° &-90° using four solar cell on each side for 100mm distance at 2mm path length) with well-known fluorescent molecules via CFIA. Iraqi Journal of Science. 2018. Vol. 59, no. 1B, pp.240-250.
https://search.emarefa.net/detail/BIM-832069
Data Type
Journal Articles
Language
English
Notes
Includes bibliographical references : p. 249-250
Record ID
BIM-832069